FUNDAMENTAL STUDIES ON ELECTRIC STIMULATOR USED FOR ACUPUNCTURE ANALGESIA
Taneomi YOSHIDAKunzo NAGAYAMAEtsutaro IKEZONOIsao MATSUMOTOTakeshi SAWA
Author information
JOURNAL OPEN ACCESS

1978 Volume 25 Issue 1 Pages 1-6

Details
Abstract
In order to perform safe and effective electric stimulation in acupuncture analgesia or therapy, basic knowledge concerning the electric stimulator is required. If direct current is used, a negative square wave with a pulse duration of 0.5~1.5 msec should be applied to the needle electrode to perform elective stimulation. Negative current is more effective at lower amperage than positive current and the possible corrosion hazard by the positive current can be avoided.