Abstract
The quantitative analysis of different calculation method of regression analysis (DCR) is performed without a calibration curve and separation process. Also, the mixture ratio of an unknown sample is determined from the coefficient of the multiple regression equation using the measured values of two or more known samples and one unknown sample. In the present work, the mixture ratio of small component in food dye mixture was calculated from 431 absorbance data of known and unknown samples measured by a spectrophotometer. As a result of calculations, a small component in a food dye mixture in an unknown sample was determined very correctly by DCR. The detection limit of the small component in the food dye mixture by this method was about 1%. The reproducibility of the purple dye mixture ratio was examined using a test sample (red dye : purple dye = 99.01 : 0.99), and the relative standard deviation from the calculated result was 3.63%. Also, the relative standard deviations of PCR and PLS were 26.36% and 12.9%, respectively. Furthermore, the existence judgment of a non-recognized small component in an unknown sample of food dye was performed based on the difference in the corrected value and the calculated value.