BUNSEKI KAGAKU
Print ISSN : 0525-1931
Technical Papers
SEM-EDX Analysis of Insulator Specimen by Using Garment Antistatic Spray
Long ZESusumu IMASHUKUJun KAWAI
Author information
JOURNAL FREE ACCESS

2013 Volume 62 Issue 2 Pages 155-158

Details
Abstract

SEM-EDX analyses of insulator samples are usually believed to be difficult because of charge-up effect, and thus carbon or gold evaporation was performed prior to the SEM observation. However, such an electric conductive thin film will emit characteristic X-rays, interfering the EDX analysis. In the present study, we report that using an anti-static spray is an easier and more effective way than using an ionic liquid. Possibility of metal corrosion by the anti-static spray should be noted.

Content from these authors
© The Japan Society for Analytical Chemistry 2013