Abstract
A method has been proposed for the gas-liquid chromatographic determination of 3, 4-dichloropropionanilide (DCPA) in technical products and emulsifiable concentrates. The DCPA was dissolved in toluene and analyzed with added benzyl benzoate (BB) as an internal standard by a Shimadzu Gas Chromatograph GC-2B with a thermal conductivity detector, using helium as carrier gas, the flow rate of which was maintained as 100ml per minute. The column (4mm×1m, stainless steel) used contained 30 weight per cent Apiezon grease L as the stationary phase and 1 weight per cent Emulsogen WS-32 as the tailing reducer on the support of Celite 545 and the column temperature was adjusted to 230°C.