Abstract
The separation efficiency of a preparative gas chromatography is often reduced by the counter-current of sample, carbonization of materials etc. Detailed examination on the temperature of the injector, amount of injection and the injecting pressure revealed that an instantaneous pressure just at the time of sample injection was quite effective to maintain good separability.