BUNSEKI KAGAKU
Print ISSN : 0525-1931
Evaluation of the mass spectrometric analysis of 13CO2 in the breath
Makiko SUEHIROMasahiro IIODale A. SCHOELLERPeter D. KLEIN
Author information
JOURNAL FREE ACCESS

1979 Volume 28 Issue 6 Pages 361-366

Details
Abstract

13CO2 in exhaled CO2 was analyzed by an isotope ratio mass spectrometer equipped with a dual inlet, a magnetic analyzer, and a single collector. Animal experiments using rats were performed for the CO2 samples. 13CO2/12CO2 in the sample was measured and 13C abundance was expressed as δ13C‰(per mil). The CO2 gas of which 13CO2/12CO2 is 0.010920 was used as the reference gas. When the sensitivity of this analysis was defined as the confidence limit in the measurement of small 13C increase, Δ13C, it was calculated to be 3.8‰. When the accuracy for the determination of Δ13C was defined as the relative deviation of the Δ13C values, it was calculated to be 1.3 ‰/Δ13C. Then if we need to obtain the accuracy with the relative deviation smaller than 5%, more than 26‰13C increase is needed. The comparison study between 14CO2 and 13CO2 analysis was also performed. Rats were given 13C and 14C labeled compounds with same labeled position simultaneously, the labeled CO2 in the breath was collected and analyzed. 14CO2 was measured by the liquid scintillation method. The recoveries of 14CO2 and 13CO2 were, as expected from the sensitivity of the 13CO2 analysis, identical when the 13C increase was more than 5‰. Comparing the sensitivity of 14CO2 and 13CO2 detection, lμCi gave the 14CO2 signals equivalent to those of 13CO2 given by 10 mM 13C. On the basis of these studies, the application of the analytical method to the medical sciences is in progress.

Content from these authors
© The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top