Flow injection determination of sulfite based on the oxidation of phosphonic acid with sulfite
Yukio HIRAINorimasa YOZAShigeru OHASHI
Author information
JOURNAL FREE ACCESS

1982 Volume 31 Issue 11 Pages 681-682

Details
Abstract
Rapid determination of sulfite was carried out by flow injection method. A sulfite solution was mixed with both a phosphonate reagent solution and an acidic molybdenum (V) -molybdenum (VI) reagent solution. In a reaction coil (PTFE, 0.5 mm i. d. 10 m, heated at 140°C), phosphonate was oxidized with sulfite and the resultant orthophosphate allowed to react with the molybdenum reagent to produce a heteropoly blue complex. The absorbance of the heteropoly blue complex was monitored at 800 nm. The standard sample could be measured in a working range of 1 × 10-4 to 1 × 10-2 M (NaHSO3) at a sampling rate of 60 samples h -1. Detection limit (S/N=2) was 2× 10-5 M. The precision (RSD, ten times injection) was 2% for sulfite of 4 × 10-3 M.