Abstract
All operations of this system are automatically and sequentially controlled by a programmer. The calibration curve is obtained by using only one standard solution. The optimal operating parameters were found for arsenic, bismuth, lead, antimony, selen, tin and mercury. The sensitiveties and detection limits of these elements are 0.15, 0.04, 0.24, 0.05, 0.05, 0.10, 0.11 and 0.1, 0.01, 0.1, 0.02, 0.04, 0.03, 0.03 ng/ml respectively. The RSDs are lower than 4% for all elements. This system was evaluated with 35 samples of various marine products, sediment and effluent.