Data Science Journal
Online ISSN : 1683-1470
Papers
Verification of the Time Accuracy of a Magnometer by Using a GPS Pulse Generator
Yasuhiro MinamotoNobuyuki KumasakaToshiaki MishimaTakeshi OowadaTakashi Okawa
Author information
JOURNAL FREE ACCESS

2011 Volume 10 Pages IAGA58-IAGA63

Details
Abstract
The time accuracy of geomagnetic data is an important specification for one-second data distributions. We tested a procedure to verify the time accuracy of a fluxgate magnetometer by using a GPS pulse generator. The magnetometer was equipped with a high time resolution (100 Hz) output, so the data delay could be checked directly. The delay detected from one-second data by a statistical method was larger than those from 0.1-s- and 0.01-s-resolution data. The test of the time accuracy revealed the larger delay and was useful for verifying the quality of the data.
Content from these authors

This article cannot obtain the latest cited-by information.

Previous article Next article
feedback
Top