GASTROENTEROLOGICAL ENDOSCOPY
Online ISSN : 1884-5738
Print ISSN : 0387-1207
ISSN-L : 0387-1207
EFFICACY OF A NEW METAL TIP CANNULA FOR SELECTIVE ERC
Naohito UCHIDAToru EZAKIShuko HIRABAYASHIAtsushi MINAMIHiroki FUKUMAKazutaka KUROKOHCHIHiroshi MATSUOKAMiho MATSUOKAMikage YACHIDAKazuhiro KOJIMAMikio NISHIOKAToshiaki NAKATSUTadashu FUJI
Author information
JOURNAL FREE ACCESS

1997 Volume 39 Issue 9 Pages 1639-1643

Details
Abstract
We evaluated the efficacy of a new metal tip ERCP cannula for patients with biliary diseases in whom cholangiography was failed. The cannula had a metal ball tip(2 mm indiarneter)with a side hole(0.6mm in diameter)deflected at 60 degree upper angulation. Of 143 patients undergoing ERCP by using a conventional cannula, cholangiography was failed in 10 cases. Employing the metal tip cannula, cholangiography was achieved in 8 of 10 failed cases(80%: new cannula versus 0%: conventional cannula: p<0.0004, Fisher's exact test). A success rate of cholangiogram was increased from 93.0% to 98.6% No complications were observed during and after the procedure. The new metal tip cannula is safe and effective for cholangiography.
Content from these authors
© Japan Gastroenterological Endoscopy Society
Previous article Next article
feedback
Top