Abstract
This paper describes an improved procedure for D/H and 18O/16O ratio analyses presented originally by Kishima and Sakai (1980). A sealed glass tube for the isotope exchange equilibration between water and CO2 gas was employed instead of a conventional container with a greased stopcock. The use of the sealed tube enabled us to handle 5 ∼ 6μmol of water for accurate and precise δ18O measurement, with precision of 0.42‰ (1 σn-1). However, for precise δD measurement more than 40μmol of water was required.