Ionizing Radiation
Online ISSN : 2758-9064
An Application of Waveform Analysis to the Pulse Signals from a 3He Gas Proportional Counter
J. KawarabayashiH. TakahashiT. IguchiM. Nakazawa
Author information
JOURNAL FREE ACCESS

1993 Volume 19 Issue 2 Pages 13-19

Details
Abstract

  With great development in the recent digital signal processing techniques, the digital waveform analysis using a fast ADC is becoming applicable to the pulse signals from radiation detectors. In the present work, this digital technique has been applied to pulse shape analysis for a 3He proportional counter. The "wall effects" in the pulse height spectrum has been successfully suppressed through detecting the pulse shape incompleteness which appears in the pulse leading edge region, when analyzing the ionization density on the triton and proton tracks produced by 3He(n,p)T reaction.

Content from these authors
© 1993 The author(s)
Previous article Next article
feedback
Top