Ionizing Radiation
Online ISSN : 2758-9064
Gamma-Ray Spectrometry with a Ge Detector: Problems of "Conventional" Energy Calibration Method
T. KishikawaK. Nishimura
Author information
JOURNAL FREE ACCESS

1993 Volume 19 Issue 3 Pages 57-71

Details
Abstract

  Problems of "conventional" energy calibration method for photon spectrometry with a germanium detector is described. Event analysis in a detection system has yielded an "instrument function" (which is the key component of a peak-shape function). The centroid of a normal distribution folded in the instrument function, instead of the position of the peak maximum (the peak position), is the intrinsic position corresponding to the reference energy. An energy (measured at the peak position of a Ge(Li) detector) of 1H(n,γ)-2224 keV is 1.3 keV smaller than that of the reference energy (2224.58900 ± 0.00022 keV).

Content from these authors
© 1993 The author(s)
Previous article
feedback
Top