2016 Volume 41 Issue 3 Pages 157-161
We fabricated a small size optical fiber type dosimeter system using Ce:CaF2 OSL element. We evaluated basic response of fabricated dosimeter to radiations of X-ray and high-energy carbon ions. When irradiating X-ray, good linearity can be confirmed. With increasing linear energy transfer (LET) of carbon ions, luminescence efficiency of Ce:CaF2 decreases. Compared with the conventional small dosimeter using Eu:BaFBr OSL element, Ce:CaF2 shows a small deterioration in the luminescence efficiency.