Japanese Heart Journal
Online ISSN : 1348-673X
Print ISSN : 0021-4868
ISSN-L : 0021-4868
Comparison of a Semi-Automated Sphygmomanometer with the Clinical Sphygmomanometer
Muhammad ILYASMasood Anwar PERACHAInayatur RAHMAN
Author information
JOURNAL FREE ACCESS

1975 Volume 16 Issue 2 Pages 118-121

Details
Abstract

A semi-automated sphygmomanometer Stanley B-200 has been compared with the clinical sphygmomanometer in 500 college boys by using a blind simultaneous method. Stanley B-200 has shown significantly higher systolic blood pressure (14.8mmHg, P<0.05) and diastolic blood pressure (3.8mmHg, P<0.05). This device also recorded 15 false positive systolic and 12 false negative diastolic results (P<0.05).