Development of a Temporal Subtraction Method with Thoracic Multi Slice CT by Use of a 3-D Image Matching
Yoshinori ITAIHyoungseop KIMSeiji ISHIKAWAShigehiko KATSURAGAWATakayuki ISHIDAIkuo KAWASHITAKazuo AWAI
Author information
JOURNAL FREE ACCESS

2007 Volume 36 Issue 6 Pages 883-890

Details
Abstract

A temporal subtraction technique can be used for enhancing interval changes on medical images by removing most the normal structures. In this study, we developed a temporal subtraction technique with multiple-detector CT (MDCT). This temporal subtraction method consisted of registration by use of a 3-dimensional image matching technique to reduce subtraction artifacts on the temporal subtraction image. Our method was applied onto 20 clinical cases including abnormal and normal cases. In the abnormal cases, at least one of abnormalities is exist on current or/and previous image. With our method, the sum and the full width at half maximum of the histogram from the temporal subtraction image, which was defined as the amount and the contrast of subtraction artifacts, were reduced by 18.9% and by 16.6%, respectively.