International Journal of Environmental and Rural Development
Online ISSN : 2433-3700
Print ISSN : 2185-159X
ISSN-L : 2185-159X
Gas Chromatography with Vacuum-Assisted Sampling System in the Analysis of Hydrogen-Balanced Nitrogen Gas
SOMPORN PLEANJAIYONGYUTH SEANSUPHANSINGTO SAKULKHAEMARUETHAIWANNASIRI SREVARITCHAWISA PINSUWANTHIDARUT LAOCHAI
Author information
JOURNAL FREE ACCESS

2014 Volume 5 Issue 2 Pages 47-50

Details
Abstract

A vacuum-assisted (VA) sampling system in gas chromatography (GC) with a thermal conductivity detector has been developed to analyze low-pressure hydrogen-balanced nitrogen gas (H2/N2) sample. The objective is to determine hydrogen-balanced nitrogen gas by GC assembled with the vacuum-assisted (VA) gas sampling system comparison to the conventional system, pressure reduced to zero (PRZ) gas sampling system. The results show that both systems were not significantly different at the 95% confidence level. The developed sampling system was applied to analyses the quality of the low-pressure-volume gas.

Content from these authors
© 2014 Institute of Environmental Rehabilitation and Conservation Research Center
Previous article Next article
feedback
Top