Accuracy Improvement of the XRD-Rietveld Method for the Quantification of Crystalline Phases in Iron Sintered Ores through the Correction of Micro-absorption Effects
Takayuki Harano Yu NemotoReiko MuraoMasao Kimura
Author information
JOURNAL OPEN ACCESS FULL-TEXT HTML

2020 Volume 60 Issue 12 Pages 2851-2858

Details
Abstract

The mass fraction of each crystalline phase in inorganic materials can be investigated using the Rietveld refinement of the X-ray diffraction (XRD) patterns. For quantitative analysis, differences in the values of the linear absorption coefficient, µ, among the crystalline phases must be considered when certain X-ray sources are used, because such differences often affect their mass fractions. Herein, we evaluate the effects of the differences between the Cu and Co X-rays on the mass fractions of the crystalline phases in iron sintered ores using the XRD-Rietveld method by performing two types of XRD measurements. Type 1 samples modeled materials with two different particle size combinations of α-Fe2O3 and ZnO. Type 2 samples used powder mixtures to simulate iron sintered ores composed of α-Fe2O3, and synthesized SFCA and SFCA-I in various mass fractions. Moreover, a correction method was developed using the Taylor-Matulis (TM) correction that considers the µ of each phase and the average particle diameter R of each crystalline phase determined by scanning electron microscopy with energy dispersive spectroscopy. For type 1 samples, results that were in good agreement with the initially-charged mass fractions could be obtained using the TM correction, even in the presence of significant differences in R between α-Fe2O3 and ZnO. The results for type 2 samples confirmed that quantitatively accurate mass fractions could be obtained using the TM correction with an accuracy of approximately ±3 mass% for Cu and Co sources, whereas the error was greater than ±3 mass% for Cu source when the TM correction was not applied.

1. Introduction

In the Asia-Pacific region, approximately 80 mass% of the iron processed in a blast furnace is obtained from iron sintered ores that consists primarily of α-Fe2O3, Fe3O4, and silico-ferrites of calcium and aluminum (SFCA, Ca2(Ca, Fe, Al)6(Fe, Al, Si)6O20,1) and SFCA-I ((Ca, Fe)4(Fe, Al)16O28)2)). In addition, iron sintered ores contain less than 10 mass% of silicate slag and amorphous materials. The crystalline type and amount of each phase are considered to be important factors for controlling the characteristics of iron sintered ores such as strength and reducibility.3,4) To quantify the crystalline phases, generally the area ratios of microtextures in cross-sectional optical microscopy images have been used.5) Other studies have been conducted to evaluate the mass fractions of the crystalline phases in iron sintered ores using electron probe microanalysis or scanning electron microscopy with energy dispersive spectroscopy (SEM-EDS).6,7) However, the identification of SFCA and SFCA-I by optical microscopy is challenging because both the chemical compositions and densities of these phases are quite similar. In addition, the solution ranges for their chemical compositions have not been determined, particularly for the non-equilibrium processes.

The X-ray diffraction (XRD) Rietveld method8) has been widely used for the quantitative analysis of crystal structures and their mass fractions in crystalline materials.9) The Rietveld method reproduces the experimental diffraction pattern for powder mixtures by simulation based on the crystallographic parameters of each phase such as the unit-cell type, atomic species, and atomic coordinates. These are used to mathematically determine the diffraction peak intensities and shapes. It is also possible to optimize the lattice parameters, the atomic occupancy of each site, and the mass fractions of the crystal phases in the samples using the non-linear least squares fitting method.8,9) In the Rietveld refinement of XRD patterns, in some cases, it is noted that each crystalline phase cannot be accurately quantified because of the particular combination of the samples and X-ray sources. Free software (RIETAN-FP,10) Profex,11) MAUD12)) as well as commercial software (PDXL2, HighScore®,13) Siroquant14)) packages have been widely used to analyze X-ray and neutron diffraction patterns.

The XRD-Rietveld method has been applied to iron sintered ores.15,16) However, in this case, particular attention must be paid to the following factors: (a) Fe, which is the main component of the ores and iron sintered ores, absorbs Cu X-rays that are often used as a X-ray source for XRD. (b) Iron sintered ores are composed of crystalline phases with a wide range of particle diameter sizes. Therefore, micro-absorption (MA) effects17) must be considered in the determination of the mass fractions of the different crystalline-phase components using the XRD-Rietveld method.

Generally, under X-ray irradiation, a portion of the X-rays are absorbed by the sample.18) The linear absorption coefficient μ that represents the degree of absorption depends on the product of the mass absorption coefficient μm and density ρ. These are determined by the atomic composition of each crystalline phase and the wavelengths (photon energy) of the incident X-rays. Furthermore, in the case of multi-phase powder samples, the XRD peak intensities of each crystalline phase varies depending on the particle diameter and packing fraction in the sample. This effect is known as the MA effect and can give rise to mass fraction errors. Brindley’s correction19) and a more advanced method (Taylor-Matulis (TM) correction20)) were previously proposed to correct the MA effect, and are expected to reduce the mass fraction error and improve the accuracy of the mass fractions of the crystalline phases in inorganic materials. Iron sintered ores consist of multiple crystalline phases that have similar crystal structures, a complicated solution with a range of cation sites, and some low-crystallinity phases. These may also affect the quantitative accuracy of the Rietveld refinement and the effect of the TM correction. A detailed discussion and quantitative analysis of the MA effects have been reported previously.21,22,23,24)

However, few studies have been conducted on the evaluation of the MA affect for the anticipated application of Rietveld analysis to iron sintered ores and iron ores. B. M. Pederson et al. evaluated the usefulness of the Brindley correction using commercial reagents.25) However, it is difficult to treat the linear absorption coefficient and the particle diameter of each crystal phase individually for the raw materials of steel, which may consist of phases with various degrees of milling as well as amorphous phases produced during the preparation of the samples for diffraction measurements. Therefore, a unified correction factor of the MA effect should not be applied to all crystalline phases in the sample. In addition, for iron sintered ores, there is an additional problem that this material tends to produce a large MA effect. This is because the constituent phases of the ores have relatively larger X-ray absorption contrast and varied particle sizes due to the hardness differences among the phases. Therefore, the degree of the MA effect and the utility of the TM correction20) are important motivations for the evaluation of the raw materials in iron-making process.

In laboratory investigations, Cu or Co radiation sources with the 1 energies of 8.048 and 6.924 keV, respectively, have been generally used as X-ray sources for powder XRD. For XRD using Cu radiation, the absorption coefficient of Fe is very large immediately above the Fe K absorption edge (7.120 keV) so that the diffraction patterns of the phases with high Fe content do not reflect real mass fractions. For an accurate quantification of the individual phases in a sample, a combination of the X-ray source and crystalline phase in the sample should be considered. Traditionally, Cu X-ray sources have been widely used in the metal industries because of their long lifetime, low cost, and high X-ray power. Therefore, in some cases, XRD measurements must be conducted using Cu radiation even though iron sintered ores contain large amounts of Fe. In other words, it is also assumed that a specific X-ray source must be used due to equipment constraints.

In this study, to improve the accuracy of the mass fractions of the individual crystal phases in iron sintered ores by the XRD-Rietveld method, the correction effect was verified for two types of samples. The first type of samples (type 1) is the mixture samples of α-Fe2O3 and ZnO with different particle sizes. In each sample, the mass fraction of each crystalline phase was calculated by the Rietveld refinement to evaluate the usefulness of the TM correction for samples with different particle sizes. The other types of sample (type 2) are mixtures of simulated iron sintered ores that consist of α-Fe2O3, SFCA, and SFCA-I. To evaluate the influence of the X-ray radiation sources not only Cu (expected large MA effect) but also Co (expected small MA effect), as general X-ray sources for XRD in laboratory studies, XRD measurements were performed to determine the quantitative values of the individual crystal phases in the samples.

2. Experimental Methods

2.1. Sample Preparation

To verify that the mass fraction accuracy of the XRD-Rietveld method can be improved by the TM corrections even when the MA effects are large, (1) sample mixtures of α-Fe2O3 and ZnO with different particle diameters and (2) sample mixtures composed of phases with different iron contents (α-Fe2O3, SFCA, SFCA-I) were prepared.

Sample mixtures of α-Fe2O3 and ZnO with different particle diameters (type 1 samples) were used for the evaluation of the MA effect and the effect of the correction for a typical sample that consists of a phase (α-Fe2O3) with a large particle diameter and linear absorption coefficient and another phase (ZnO) with small particle diameter and linear absorption coefficient. Powder samples in which α-Fe2O3 and ZnO, with the average particle diameters of 0.5 or 20 μm and 0.6 μm, respectively, were mixed in an equal mass fraction (50%/50%). Samples with the same mass fraction mixtures of α-Fe2O3 and ZnO were prepared by mixing reagent grade α-Fe2O3 (Kojundo Chemical Laboratory Co., Ltd., Japan, 4N, 0.5 μm) or α-Fe2O3 with a 20-μm average particle size (measured by Laser scattering measurement) obtained by grinding 2-mm particles of α-Fe2O3 (Kojundo Chemical Laboratory Co., Ltd., Japan, 4N), and ZnO (Kojundo Chemical Laboratory Co., Ltd., Japan, 4N, 0.6 μm) in equal mass (50%/50%).

Sample mixtures composed of phases with different iron contents (α-Fe2O3, SFCA, SFCA-I) (type 2 samples) were used to evaluate the accuracy of the mass fraction of each phase in the simulated iron sintered ores for the correction. Six model samples consisting of three compounds were prepared. Specifically, α-Fe2O3, which is the main phase in iron sintered ores, and two types of synthesized silico-ferrites of calcium and aluminum (SFCA, SFCA-I), that have particularly unique chemical compositions,1,2) were mixed in various mass fractions. Samples of SFCA and SFCA-I were synthesized by conventional powder sintering following the procedures reported by J. D. I. Hamilton et al.1) and W. G. Mumme et al.2) Their detailed crystal structures including atomic coordinates were reported in the literatures.1,2) For the synthesis of single-phase SFCA and SFCA-I, reagent grade α-Fe2O3 (Kojundo Chemical Laboratory Co., Ltd., Japan, 4N, 0.5 μm), CaCO3 (Kanto Chemical Co., Ltd., Japan, 4N, 12 μm), α-SiO2 (Kojundo Chemical Laboratory Co., Ltd., Japan, 99%, 1 μm), α-Al2O3 (Kojundo Chemical Laboratory Co., Ltd., Japan, 4N, 1 μm), and MgO (Kojundo Chemical Laboratory Co., Ltd., Japan 4N, 0.6 μm) were used. Powder mixtures (approximately 3 g) prepared with the composition shown in Table 1 were pelletized into two or three tablets with a diameter of 13 mm at a pressure of 34.5 kN and then were calcined at 800°C. Annealing was performed at 1200°C (SFCA) and 1230°C (SFCA-I) for 60 h, respectively. The tablets were then pulverized and pelletized into new tablets with the same size, and the same heat treatment was repeated three times. Figure 1 shows the XRD patterns (Cu ) of the synthesized SFCA and SFCA-I. Although α-Fe2O3 was included in the synthesized SFCA and SFCA-I samples, single phases with a purity of greater than 99% and 95% were obtained, respectively, as deduced from the XRD peak intensities of each sample. Synthesized SFCA, SFCA-I, and α-Fe2O3 were mixed at the mass fractions listed in Table 2 to prepare model samples. α-Fe2O3 was prepared by pulverizing a reagent grade shot (Kojundo Chemical Laboratory Co., Ltd., Japan 4N) with an average particle size of approximately 2 mm. After blending, each model sample was pulverized and mixed for approximately 30 min using an agate mortar and a pestle to ensure homogeneity.

Table 1. Initial compositions of SFCA and SFCA-I (mass%).
CaCO3α-Fe2O3α-Al2O3α-SiO2MgO
SFCA1)25.759.15.37.82.1
SFCA-I2)18.276.94.9
Fig. 1.

XRD patterns (Cu ) of synthesized SFCA (red) and SFCA-I (blue). The rows of the bars indicate the positions of the XRD peaks of SFCA, SFCA-I, and α-Fe2O3 reported in previous studies1,2,26) from the top.

Table 2. Mass fractions of model samples (mass%).
No.α-Fe2O3SFCASFCA-I
150.025.025.0
249.812.637.6
349.837.512.7
450.033.416.6
550.039.810.3
649.945.15.0

2.2. XRD Measurements

To verify that the accuracy of the XRD-Rietveld method can be improved by the TM corrections even when the MA effects are large, two cases are considered; (1) mixture samples of α-Fe2O3 and ZnO with different particle sizes using X-ray sources of Cu and (2) samples simulated iron sintered ores using X-ray sources of Cu and Co .

X-rays of Cu radiation were used for the XRD measurements performed to verify the mass fraction accuracy of the TM correction, while in the XRD measurements performed to examine the effect of X-ray sources, the XRD patterns were obtained using X-rays of both Cu and Co sources. XRD patterns were analyzed by the Rietveld method to investigate the mass fractions of the crystalline phases. In addition, we attempted to correct each quantitative mass fraction using the correction of J. C. Taylor et al.20) in which we considered the linear absorption coefficient and average particle diameter of each phase. In type 2 samples, the average particle diameters of α-Fe2O3, SFCA, SFCA-I in each sample powder obtained by SEM-EDS with automatic particle analysis were also used for the correction.

A Smartlab diffractometer (Rigaku, Japan) was used for XRD measurements. Powders of model samples were filled in a sample holder (sample part: ϕ 20 mm × 0.2 mm depth), and XRD measurements were performed. For Cu radiation, the tube current and tube voltage were set at 40 mA and 40 kV, respectively. For Co radiation, the tube current and tube voltage were set at 36 mA and 40 kV, respectively. The goniometer radius was 285 mm. A high-speed, one-dimensional detector D/teX (Rigaku, Japan) was used as the X-ray detector, and measurements were performed using the Bragg-Brentano focusing geometry (i.e., the filter method, with Cu and Co using Ni and Fe filters, respectively). The measurement conditions were as follows: Measurement angle range 2θ = 20–80°, step angle width Δ2θ = 0.02°, scan speed = 1°/min, and sample azimuth rotation speed = 60 rpm. The divergence slit (DS) was 1/2°, and the widths of the receiving slits RS1 and RS2 were 8 and 13 mm, respectively.

To carry out the XRD-Rietveld analysis, the PDXL2 software (version 2.8.1.1, Rigaku, Japan) was used. The initial crystallographic parameters of each crystalline phase used in the analysis were as follows: ICDD card number 01-080-237726) for α-Fe2O3, ICDD card number 01-079-020627) for ZnO, parameters reported by J. D. I. Hamilton1) for SFCA, and parameters reported by W. G. Mumme2) for SFCA-I. In the analysis, the atomic coordinates and temperature factors were fixed at the initial values. The full width at half maximum parameters (U, V, W ) of the profile function for the samples of both types 1 and 2 of α-Fe2O3 and those of the other crystal phases were set to the same values, and non-linear least square fitting was performed. The Rietveld method is described in detail in the literature.10)

2.3. Taylor-Matulis Correction

According to J. C. Taylor et al.,20) the correction factor τi for the MA effect in XRD patterns for the quantitative value of the ith crystal phase can be expressed by Eq. (1):   

τ i = 1 A i 0 A i e -( μ i - μ ¯ ) x d A i (1)
where Ai is the particle volume, μi is the linear absorption coefficient of the ith crystal phase, and μ is the average linear absorption coefficient of the samples. μ is calculated from the mass fraction obtained prior to the TM correction. Assuming that the particle is a perfect sphere, Eq. (1) can be simplified, yielding the correction coefficient τi given by Eq. (2):   
τ i = 3 X 3 ( e X ( X 2 -2X+2 ) -2 ) (2)
where X = (μiμ)Ri. and Ri is the average diameter of the ith crystalline phase. When the correction coefficient τi is calculated in this manner, the mass fraction of the ith crystalline phase is expressed as shown in Eq. (3),   
w i = s i Z i M i V i τ i j=1 n s j Z j M j V j τ j = w i τ i j=1 n w j τ j (3)
where si, Zi, Mi, Vi, wi, and w i are the scale factor in the XRD-Rietveld method, the number of molecules per unit cell, the formula weight, unit cell volume, and the mass fraction after and before correction of the ith crystal phase, respectively. Here, n is the number of the crystal phases in the samples.

2.4. SEM-EDS Measurements

SEM-EDS measurements were performed to determine the average particle diameter of α-Fe2O3 and the multi-component calcium ferrite included in sample No. 1 with the mass fractions shown in Table 2. Metal Quality Analyzer (MQA, Thermo Fisher Scientific, USA) was used for the SEM-EDS measurements. The sample was prepared as follows. An aqueous solution in which the powder sample No. 1 was prepared (as described in Section 2.1) was dispersed and collected with a dropper. Dispersion was achieved by dropping the sample particles on a 10 × 10 × t1 mm3 Si substrate with a carbon tape attached. After drying, the particle diameters and compositions of approximately 7000 particles were analyzed at approximately 3000 particles/h using an automatic particle analysis function of MQA. In this study, the particles in which only Fe and O was detected by EDS were identified as α-Fe2O3, and the particles in which Al, Ca, Fe, O, and Si were detected were identified as multi-component calcium ferrite (SFCA + SFCA-I) because SFCA and SFCA-I that are found in different solid solution states cannot be distinguished in iron sintered ores in real steel works. In addition, each particle size distribution was measured simultaneously. Figure 2 shows the distributions of the particle diameters of α-Fe2O3 and calcium ferrite (SFCA + SFCA-I). The average particle diameters of α-Fe2O3 and calcium ferrite were defined as the arithmetic mean of the particles measured in the two cardinal and two ordinal directions.

Fig. 2.

Distributions of the average particle diameter sizes of α-Fe2O3 and multi-component calcium ferrite (SFCA+SFCA-I) obtained by SEM-EDS measurements.

3. Results and Discussion

To verify that the accuracy of the XRD-Rietveld method can be improved by the TM corrections even when MA effects are large, two cases are considered; (1) sample mixtures of α-Fe2O3 and ZnO particles with different diameters using X-ray sources of Cu and (2) samples simulated iron sintered ores using X-ray sources of Cu and Co .

3.1. Validation of the Taylor-Matulis Correction for the Micro-absorption Effects

Figure 3 shows the XRD patterns of α-Fe2O3 and ZnO in the 1:1 mass ratio obtained using a Cu X-ray source. The blue line shows the pattern of the mixture of α-Fe2O3 and ZnO with the average particle sizes of 0.5 and 0.6 μm, respectively. The red line shows the pattern of the mixture of α-Fe2O3 and ZnO with average particle sizes of 20 and 0.6 μm, respectively. For the powder sample with the average particle diameters of 20 μm for α-Fe2O3 and 0.6 μm for ZnO, the quantitative values (α-Fe2O3: 35 mass%, ZnO: 65 mass%) deviated significantly from the initially-charged mass fraction. In the Rietveld refinement, the corresponding Rwp/S factors were 3.42%/2.52 and 2.56%/1.91, respectively. Note that Rwp and S are the weighted profile R-factor and the goodness-of-fit, respectively. The correction coefficient τ was calculated by the TM correction using the linear absorption coefficients at the X-ray energy of Cu 1 and the average particle diameters of α-Fe2O3 and ZnO. The linear absorption coefficients and particle diameters of α-Fe2O3 and ZnO, as well as τFe2O3 and τznO are presented in Table 3. It should be noted that τFe2O3 and τznO were calculated using the initially-charged mass fraction of each sample, because the goal of the experiments on type 1 samples was to examine the accuracy and usefulness of the TM correction.

Fig. 3.

XRD patterns of α-Fe2O3 and ZnO in the 1:1 mass ratio measured using Cu radiation. Dotted blue and red lines show the diffraction patterns for the samples where the average particle sizes of α-Fe2O3 and ZnO are 0.5 and 0.6 μm, and 20 and 0.6 μm, respectively. The inset shows the region where the difference is prominent. The Rwp/S factors of the Rietveld analysis are 3.42%/2.52 and 2.56%/1.91, respectively. The rows of the bars indicate the positions of the XRD peaks of α-Fe2O3 and ZnO reported in previous studies26,27) from the top.

Table 3. MA correction parameters for α-Fe2O3 and ZnO mixture specimens.
SpeciesLinear absorption coefficient μi (cm-1)Averaged particle diameter R (μm)μ (cm−1)Correction factor τ
α-Fe2O311190.5689.50.984
111920689.50.533
ZnO2600.6689.51.02

The quantitative mass fractions before and after the application of the TM correction of the XRD pattern for the samples with the average particle sizes of α-Fe2O3 and ZnO of 0.5 and 0.6 μm, respectively, are depicted in Fig. 4 as A and A’, respectively. Both sets of mass fractions results were almost identical to each other because the measurement conditions were ideal in terms of the average particle diameters of α-Fe2O3 and ZnO. For the XRD patterns of the samples with the average particle diameters of α-Fe2O3 and ZnO of 20 and 0.6 μm, the quantitative mass fractions before and after the application of the TM correction are depicted as B and B’ in Fig. 4, respectively. Although the B results could not reproduce the mixing mass fractions, it was confirmed through the calculated correction factors τFe2O3 and τznO that the quantitative values of B’ as shown in Fig. 4 were nearly the same as the initially-charged composition, even when the particle diameter difference was large. The penetration depth tp is given as μtp = 1, and tp = 18 and 70 μm for α-Fe2O3 and ZnO, respectively when Cu is used. On the other hand, tp = 89 and 46 μm for α-Fe2O3 and ZnO, respectively when Co is used. Therefore, these results show that it is particularly important to consider the MA effects when the particle diameter of the phase that absorbs the X-ray is as large as the penetration depth tp.

Fig. 4.

Mass fractions before and after the Taylor-Matulis correction determined by the Rietveld analysis of each XRD pattern shown in Fig. 3. A and A’ indicate the quantitative values before and after the Taylor-Matulis correction of the XRD patterns with the average α-Fe2O3 and ZnO particle sizes of 0.5 and 0.6 μm, respectively. B and B’ indicate the quantitative values before and after the correction of the XRD patterns for α-Fe2O3 and ZnO with the average particle diameter sizes of 20 and 0.6 μm, respectively.

It was confirmed that the accuracy of the XRD-Rietveld method can be improved by the TM correction through the comparison of the results for the mixture samples of α-Fe2O3 and ZnO with different particle diameters, namely (A) 0.5 and 0.6 μm, and (B) 20 and 0.6 μm, respectively.

3.2. The Taylor-Matulis Correction of the Difference in the X-ray Sources in Samples Simulated Iron Sintered Ores

We next considered the case of type 2 samples. Table 4 shows the average particle diameters as described in Section 2.4, the linear absorption coefficients of Cu 1 and Co 1, and the τ factors of α-Fe2O3, SFCA, and SFCA-I for each model sample. The average particle diameters of α-Fe2O3 and multi-component calcium ferrite were 4.1 and 4.6 μm, respectively, showing that a correction of 5–8% is required for Cu . The grain sizes are smaller than the values of the penetration depth tp, and the MA effects are not as strong as in type 1 samples.

Table 4. Correction factors, τ, for α-Fe2O3 and silico-ferrite of calcium and aluminum (SFCA+SFCA-I).
speciesμ (Cu 1) (cm−1)μ (Co 1) (cm−1)Averaged particle diameter R (μm)No.μ (cm−1)Correction factor τ for Cu 1Correction factor τ for Co 1
α-Fe2O311192254.11945.70.9481.006
2961.20.9531.007
3929.10.9431.006
4934.90.9451.006
5926.60.9431.006
6919.60.9411.006
SFCA7082664.61945.71.0860.993
2961.21.0920.993
3929.11.0790.993
4934.91.0820.993
5926.61.0790.993
6919.61.0760.993
SFCA-I8372684.61945.71.0380.992
2961.21.0440.993
3929.11.0320.992
4934.91.0340.992
5926.61.0310.992
6919.61.0290.992

In the following analysis, the TM correction calculations were performed for all samples using the average particle diameters of α-Fe2O3, SFCA, and SFCA-I in sample No. 1. The particle diameters of each crystalline phase in all samples can be considered to be approximately the same because all of the samples were prepared and mixed using the same method. The linear absorption coefficients of SFCA and SFCA-I were calculated using the compositions and density reported in the literature.1,2)

Figure 5 shows the XRD patterns of the model samples (Nos. 1–6) consisting of α-Fe2O3, SFCA, and SFCA-I1,2,26) using Cu radiation. In the bottom part of Fig. 5, the XRD patterns of α-Fe2O3, SFCA, and SFCA-I reported in previous studies are shown.1,2,26) The rows of the bars under patterns indicate the peak positions of α-Fe2O3, SFCA, and SFCA-I from the top. Figure 6 shows an XRD pattern (red) of the model sample (No. 1) obtained using Cu , the simulated pattern (blue) obtained by the Rietveld method, and the residual pattern (pink) obtained as the difference between the red and blue curves. Figures 7 and 8 show the results corresponding to those presented in Figs. 5 and 6, respectively, but obtained using Co source. Figures 5 and 7 show that the intensity of the peaks near 2θ = 34.5° and 40.5° changed due to the difference in the mass fraction between SFCA and SFCA-I, respectively. It was also confirmed that the measured XRD patterns presented in Figs. 6 and 8 could be reproduced well by the Rietveld refinement. Table 5 shows the Rwp and S factors of the Rietveld refinements of the XRD patterns shown in Figs. 5 and 7.

Fig. 5.

(a) XRD patterns of the model samples (No. 1–6) consisting of α-Fe2O3, SFCA, and SFCA-I using Cu . (b) Magnified view of a region of Fig. 5(a). The rows of the bars indicate the positions of XRD peaks of α-Fe2O3, SFCA, and SFCA-I reported in previous studies1,2,26) from the top.

Fig. 6.

XRD pattern (red) of the model sample (No. 1) obtained using Cu radiation and the simulated pattern (blue) obtained by the Rietveld method, and the residual pattern (pink) showing the difference between the red and blue curves. The background pattern is shown by the green curve.

Fig. 7.

XRD patterns of the model samples (No. 1–6). XRD patterns of the samples consisting of α-Fe2O3, SFCA, and SFCA-I obtained using Co radiation. The rows of bars under patterns indicate peak positions of α-Fe2O3, SFCA, and SFCA-I reported in previous studies1,2,26) from the top.

Fig. 8.

XRD pattern (red) of the model sample (No. 1) obtained using Co radiation and the simulated pattern (blue) obtained by the Rietveld method, and residual pattern (pink) showing the difference between the red and blue curves. The background pattern is shown by the green curve.

Table 5. Rwp and S factors of the Rietveld refinement of XRD patterns shown in Figs. 5 and 7.
No.Cu Co
RwpSRwpS
11.941.461.471.60
22.031.551.341.47
32.131.611.511.63
42.031.551.551.68
52.251.701.681.82
62.261.701.651.79

Figure 9 shows the relationship between the initially-charged mass fraction of each sample (shown in Table 2) and the mass fractions obtained by the XRD-Rietveld method before and after the TM correction using Eq. (3) and their correction factors presented in Table 4. Figures 9(a) and 9(b) show the results for Cu 1 and Co 1 radiation, respectively. Table 6 shows the data presented in Figs. 9(a) and 9(b). The dotted lines in Figs. 9(a) and 9(b) represent a deviation of the quantitative mass fractions of ±3 mass%. The results obtained using Cu radiation without the correction of the MA effect reveal that the mass fraction of α-Fe2O3, with the largest linear absorption coefficient of 1119 cm-1 was smaller than the initially-charged mass fraction. However, each mass fraction of silico-ferrite of calcium and aluminum (SFCA, SFCA-I) with a relatively small linear absorption coefficient was greater than the initially-charged mass fraction. The TM correction also enabled us to obtain quantitative mass fractions that were nearly identical to the initially-charged mass fractions. Furthermore, for the Co data, the mass fraction without the correction was nearly the same as the initially-charged mass fraction. Practically, no correction effect was observed, and most samples could be quantified within ±3 mass% because the linear absorption coefficients were small for all phases. These results confirmed that the crystalline phase in samples can be quantified with accuracy of ±3 mass% using the correction method developed in this study, even for measurements with Cu radiation. Thus, MA affects the mass fraction of each phase but has little effect on the shape of the XRD pattern. This may be because the specimens are polycrystalline and enough fine, where a sufficiently large number of grains are observed by XRD in the cases of both small and large MA effects. This accuracy is sufficient for the use of the method for evaluation in the iron-making process and for the determination of the relationship between the mass fractions and qualities of iron sintered ores. However, two additional points should be considered in the application of this technique to iron sintered ores in the industrial process. The first is the low crystallinity values of SFCA and SFCA-I in a real iron sintered ores. The second is the variation of the chemical composition in SFCA and SFCA-I that also broadens the XRD peak widths. These points make XRD peaks of SFCA and SFCA-I broader and it may be difficult to fit properly them reflecting to the mass fractions of SFCA and SFCA-I in iron sintered ores.

Fig. 9.

Relationship between the initially-charged mass fractions of the samples (shown in Table 2) and the quantitative values obtained by the XRD-Rietveld method before and after the TM correction using Eq. (3). Figures 9(a) and 9(b) show the results obtained for the Cu 1 and Co 1 radiation sources, respectively.

Table 6. Mass fraction of each crystalline phase before and after the correction in XRD measurements of type 2 samples (α-Fe2O3: H, SFCA: S0, SFCA-I: S1).
Cu Co
Initially-charged mass fractionBefore correctionAfter correctionBefore correctionAfter correction
HS0S1HS0S1HS0S1HS0S1HS0S1
No. 150.025.025.045.928.026.148.726.025.351.724.024.351.424.224.5
No. 249.812.637.645.216.538.447.715.237.051.513.535.051.213.635.2
No. 349.837.512.744.441.314.347.538.614.052.235.212.651.835.512.7
No. 450.033.416.646.935.118.049.932.617.553.530.316.253.030.616.4
No. 550.039.810.344.743.811.547.841.011.251.837.610.651.537.810.7
No. 649.945.15.046.648.25.249.845.15.153.341.84.853.042.14.9

Based on the above-described results, for the type 2 samples, it was confirmed that the accuracy of the XRD-Rietveld method can be improved by the TM correction even for Cu X-ray radiation for which the XRD-Rietveld method may not be able to determine the mass fractions with quantitative accuracy due to the MA effect.

4. Conclusion

We investigated the effect of the TM correction on the mass fractions of the crystalline phases of mixtures of α-Fe2O3 and ZnO (type 1) and the effects of the X-ray source and the TM correction on the mass fractions of a simulated iron sintered ores samples (type 2) using the XRD-Rietveld method in experiments aimed at the application of this method in iron-making process.

For Cu radiation with the energy close to that of the Fe K-edge absorption edge, we confirmed that the mass fractions were determined in accordance with the linear absorption coefficient of the constituent phases. In type 1 samples, when both particle sizes were sufficiently small, quantitatively accurate mass fractions were obtained as shown by the comparison to the initially-charged mass fractions. However, when the difference in particle diameters was large, quantitatively accurate mass fractions reflecting the initially-charged mass fractions of the individual crystalline phase could not be obtained. However, the mass fractions for the mixing sample were obtained using the TM correction that considers the linear absorption coefficient calculated from the chemical composition and density.

For type 2 samples, the correction method in advance of the TM correction for obtaining quantitatively accurate mass fractions of the simulated iron sintered ores was developed by considering the average particle diameter and linear absorption coefficient of the components obtained by SEM-EDS (MQA) measurements through automatic particle analysis. With this correction method, the quantitative mass fractions could be obtained to an accuracy of ±3 mass% even when Cu radiation was used for samples that included iron oxides. In addition, we confirmed that the mass fraction of each crystalline phase could be quantified with accuracy of ±3 mass% or less even without the application of the correction method when Co radiation was used. This was due to the small difference in the linear absorption coefficient between the Fe-based oxide and silico-ferrite of calcium and aluminum. Thus, our results show that even when using an X-ray source that was strongly absorbed by the samples, the correction method developed in this study can provide more accurate quantitative mass fractions of the individual crystal phases. It is expected that this method can be applied not only to model samples but also for industrial materials.

Acknowledgments

We thank Dr. Takafumi Takahshi, Dr. Koji Kanehashi, Dr. Kyosuke Hara, Dr. Kenichi Higuchi, Mr. Toru Takayama, Dr. Michihiro Aimoto, and Dr. Koji Saito from the Nippon Steel Corporation, Dr. Jun Okazaki from the Nippon Steel Technology Co. Ltd., Prof. Takahiro Sagawa from the Maebashi Institute of Technology, and Prof. Takashi Ida from the Nagoya Institute of Technology for useful discussions about the XRD-Rietveld method and iron sintered ores. We are also grateful to Dr. Daisuke Itabashi from the Nippon Steel Corporation for technical advice and support in conducting the SEM-EDS measurements.