LASER THERAPY
Online ISSN : 1884-7269
Print ISSN : 0898-5901
ISSN-L : 0898-5901
Original Article
Influence of diffuse reflectance measurement accuracy on the scattering coefficient in determination of optical properties with integrating sphere optics (a secondary publication)
Takuro HoribeKatsunori IshiiDaichi FukutomiKunio Awazu
Author information
JOURNAL FREE ACCESS

2015 Volume 24 Issue 4 Pages 303-310

Details
Abstract

An estimation error of the scattering coefficient of hemoglobin in the high absorption wavelength range has been observed in optical property calculations of blood-rich tissues. In this study, the relationship between the accuracy of diffuse reflectance measurement in the integrating sphere and calculated scattering coefficient was evaluated with a system to calculate optical properties combined with an integrating sphere setup and the inverse Monte Carlo simulation. Diffuse reflectance was measured with the integrating sphere using a small incident port diameter and optical properties were calculated. As a result, the estimation error of the scattering coefficient was improved by accurate measurement of diffuse reflectance. In the high absorption wavelength range, the accuracy of diffuse reflectance measurement has an effect on the calculated scattering coefficient.

Content from these authors
© 2015 Japan Medical Laser Laboratory
Previous article Next article
feedback