A method for measuring the longitudinal chromatic aberration of the human eye with a Shack-Hartmann wavefront sensor
Masashi NAKAJIMAYoko HIROHARAMariko KOBAYASHIToshifumi MIHASHI
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Details
Abstract
We developed an instrument for measuring the longitudinal chromatic aberration (LCA) in three different wavelengths (561, 690, 840 nm) using Shack-Hartmann wavefront aberrometer. Switching three light sources consecutively, we can measure the LCAs of the human eyes precisely. The wavefront aberration is calibrated by using a reflecting model eye which has no chromatic aberration in principle and by using a ISO model eyes, of which lens designs were known. We confirmed that the longitudinal chromatic aberration of the human eyes can be measured with this method.