Journal of the Atomic Energy Society of Japan / Atomic Energy Society of Japan
Online ISSN : 2186-5256
Print ISSN : 0004-7120
ISSN-L : 0004-7120
Remotely Operated Pipetter
Satoru EMURAYoshinori NAKAHARAMamoru ONUKINobuaki KONO
Author information
JOURNAL FREE ACCESS

1967 Volume 9 Issue 12 Pages 705-709

Details
Abstract
Two positive delivery pipetters (1 ml and 5ml in capacity) have been devised for pipetting aliquots of highly radioactive liquid samples intended, various analytical deteminations in the reprocessing analytical facility at the Japan Atomic Energy Research Institute.
The pipetter assembly consists of pipet, electrical elevator, control unit, three valves and trap, as well as nitrogen gas and vacuum lines.
The structural material of the parts of the pipet in contact with the sample are either stainless steel or teflon.
A precision of 0.2% standard deviation of successive 100μl deliveries of distilled water and concentrated nitric acid was obtained on the 1 ml pipetter.
Design criteria for the pipetter, its principle of operation and method of decontamination are described.