2018 Volume 67 Issue 1 Pages 52-58
Insulin testing is widely used in the diagnosis and assessment of the stage of glucose metabolism disorders. In this study, we evaluated the performance characteristics and clinical utility of “Norudia® Insulin”. The coefficients of variation (C.V.) for within-run and between-run precisions were 1.25–3.08% and 3.69–8.07%, respectively. Linearity was maintained up to 144.2 μU/mL, and the quantitation limit was 1.0 μU/mL. Prozone effects were observed at ≥ 202.5 μU/mL. There was significant interference by hemoglobin. Comparison with “ARCHITECT Insulin” showed good correlation with a regression equation of y = 0.96x + 1.0 and a correlation coefficient of r = 0.992 (samples from patients not receiving insulin analog formulations). In addition, the reagent was confirmed to not react with insulin analog formulations after analytical recovery testing. The results of a questionnaire survey targeting doctors at our hospital revealed that assays that can selectively measure endogenous insulin are highly desirable; in addition, they can reduced testing times and costs. As such, the assay should be a useful tool in clinical settings.