Journal of the Anus, Rectum and Colon
Online ISSN : 2432-3853
ISSN-L : 2432-3853
Original Research Articles
Verification of the Storage Stability of Fecal Hemoglobin in Stool Sampling Bottle Using Improved Buffer - A Pilot Study -
Masakatsu Fukuzawa, Sakiko Naito, Shin Kono, Yoshiya Yamauchi, Takashi Morise, Kumiko Uchida, Yasuyuki Kagawa, Takahiro Muramatsu, Shiori Enomoto, Shizuka Takehara, Norimitsu Hosaka, Takao Itoi
Author information
JOURNAL OPEN ACCESS

2026 Volume 10 Issue 3 Pages 375-383

Details
Abstract

Objective: The temperature and duration of sample storage is known to affect the accuracy of the fecal immunochemical test (FIT). This study compared an improved stool collection buffer with the conventional buffer on the fecal hemoglobin (Hb) correlation and high-temperature stability and simultaneously evaluated the stability of fecal transferrin (Tf) and fecal calprotectin (FC).

Methods: From June 2021 to December 2024, 130 stool samples collected from patients with informed consent were analyzed. Fecal Hb, Tf, and FC were measured in samples stored in both buffers, followed by correlation and stability assessments. High-temperature stability was evaluated at 37°C for 1 week.

Results: The improved buffer showed strong correlations with the conventional buffer for three markers (r ≥0.95; slope 1.00 ± 0.10). In the high-temperature stability test, residual reactivities were assessed for Hb ≥10 μg/g, Tf ≥10 μg/g, and FC ≥50 μg/g. Hb residual reactivity was significantly higher with the improved buffer (p < 0.01), while Tf maintained high stability (> 90%) in both buffers. FC showed reduced stability in the conventional buffer under prolonged high-temperature exposure, whereas the improved buffer demonstrated better preservation. The improved buffer also maintained higher Hb stability at elevated concentrations (≥30 μg/g).

Conclusions: The improved stool collection buffer effectively addressed the limitations of the conventional buffer by providing better Hb stability under high-temperature and high-concentration conditions, while preserving excellent stability for Tf and FC. These findings support its potential utility for postal stool collection in FIT-based colorectal cancer screening.

Fullsize Image
Content from these authors
© 2026 The Japan Society of Coloproctology

JARC is an Open Access journal distributed under the Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License. Anyone may download, reuse, copy, reprint, or distribute articles published in the Journal for non-profit purposes if they cite the original authors and source properly. If anyone remixes, transforms, or builds upon the material, you may not distribute the modified material.
https://creativecommons.org/licenses/by-nc-nd/4.0/
Previous article Next article
feedback
Top