EVALUATION OF INTER-FRACTION ERROR DURING PROSTATE RADIOTHERAPY
Takafumi KOMIYAMAKoji NAKAMURATsuyoshi MOTOYAMAHiroshi ONISHINaoki SANO
Author information
JOURNAL FREE ACCESS

2008 Volume 20 Issue 1 Pages 1-6

Details
Abstract

Purpose: To evaluate inter-fraction error (inter-fraction set-up error+inter-fraction internal organ motion) between treatment planning and delivery during radiotherapy for localized prostate cancer.
Methods and Materials: Twenty three prostate cancer patients underwent image-guided radical irradiation with the CT-linac system. All patients were treated in the supine position. After set-up with external skin markers, using CT-linac system, pretherapy CT images were obtained and isocenter displacement was measured.
Results: The mean displacement of the isocenter was 1.8 mm, 3.3 mm, and 1.7 mm in the left-right, ventral-dorsal, and cranial-caudal directions, respectively. The maximum displacement of the isocenter was 7 mm, 12 mm, and 9 mm in the left-right, ventral-dorsal, and cranial-caudal directions, respectively. The mean interquartile range of displacement of the isocenter was 1.8 mm, 3.7 mm, and 2.0 mm in the left-right, ventral-dorsal, and cranial-caudal directions, respectively.
Conclusion: In radiotherapy for localized prostate cancer, inter-fraction error was largest in the ventral-dorsal directions. Errors in the ventral-dorsal directions influence both local control and late adverse effects. Our study suggested the set-up with external skin markers was not enough for radical radiotherapy for localized prostate cancer, thereby those such as a CT-linac system for correction of inter-fraction error being required. (Article in Japanese)