EVALUATION OF ISOCENTER REPRODUCIBILITY IN TELEMEDICINE OF 3D-RADIOTHERAPY TREATMENT PLANNING
Saeko HIROTAKayoko TSUJINOKouji KIMURAYoshiki TAKADAYoshio HISHIKAWAMichio KONOToshinori SOEJIMAAkihisa KODAMA
Author information
JOURNAL FREE ACCESS

2000 Volume 12 Issue 3 Pages 259-266

Details
Abstract
Purpose: To evaluate the utility in telemedicine of Three-Dimensional Radiotherapy Treatment Planning (tele-3D-RTP) and to examine the accuracy of isocenter reproducibility in its offline trial.
Materials and Methods: CT data of phantoms and patients in the satellite hospital were transferred to our hospital via floppy-disk and 3D-radiotherapy plans were generated by 3D-RTP computer in our hospital. Profile data of CT and treatment beams in the satellite hospital were pre-installed into the computer. Tele-3D-RTPs were performed in 3 phantom plans and 14 clinical plans for 13 patients. Results: Planned isocenters were well reproduced, especially in the immobilized head and neck/ brain tumor cases, whose 3D-vector of aberration was 1.96±1.38 (SD) mm.
Conclusion: This teletherapy system is well applicable for practical use and can provides cost-reduction through sharing the resources of expensive equipment and radiation oncologists.
Content from these authors
© 1994 The Japanese Society for Therapeutic Radiology and Oncology