2025 Volume 32 Issue 5 Pages 580-595
Aim: Lipoprotein (a) [Lp(a)] is a well-established risk factor for cardiovascular disease independent of low-density lipoprotein-cholesterol (LDL-C). The Lp(a) concentrations were inconsistent between the immunoassays. This study aimed to investigate whether harmonization of Lp(a) measurements can be achieved using a serum panel value assigned with the IFCC-endorsed mass spectrometry-based reference measurement procedure (IFCC-MS-RMP).
Methods: We measured the Lp(a) concentrations using five Lp(a) immunoassays in 40 panel sera provided by the Centers for Disease Control and Prevention (CDC), and 500 Japanese subjects enrolled in the Bunkyo Health Study. Of the five immunoassays, only the Roche Lp(a) assay was traceable to the WHO-IFCC reference material SRM2B. Lp(a) concentrations in CDC samples were also determined by IFCC-MS-RMP, provisionally calibrated to SRM2B. Lp(a) concentrations were expressed in mass units (mg/dL) for most reagents, but in SI units (nmol/L) for Roche’s reagent and IFCC-MS-RMP.
Results: In the CDC panel sera, all immunoassays, including Roche’s reagent, showed good correlations with IFCC-MS-RMP. In the Bunkyo Health Study samples, all immunoassays showed good correlations with Roche’s reagent (rs, 0.986-0.998) although the slopes of the regression lines ranged from 0.292 to 0.579. After recalibration with the CDC’s panel sera, Lp(a) results of Bunkyo Health Study samples were converted to the equivalent values determined by the IFCC-MS-RMP, thus resulting in a marked reduction in the intermethod CV among the assays.
Conclusion: We achieved harmonization of Lp(a) measurements with five immunoassays using a serum panel value assigned with the IFCC-MS-RMP.
See editorial vol. 32: 563-564
Numerous epidemiologic and clinical studies have consistently shown that low-density lipoprotein-cholesterol (LDL-C) is a strong risk factor for cardiovascular disease (CVD), even after adjusting for other risk factors1, 2). Although aggressive LDL-C-lowering therapy with a combination of potent statins and proprotein convertase subtilisin/kexin type 9 (PCSK9) inhibitors can significantly reduce CVD in high-risk patients, we observed persistent residual CVD risk in these patients3). Among other lipoprotein-related biomarkers, many epidemiological and clinical studies have shown that lipoprotein (a) [Lp(a)] is an independent risk factor for CVD4, 5) and aortic valve calcification/stenosis6). Furthermore, the association between Lp(a) and CVD has been confirmed by recent Mendelian randomization (MR) studies using several Lp(a)-increasing single nucleotide polymorphisms (SNPs)7). These data strongly suggest a causal role for Lp(a) in the development of CVD. Despite the impact of Lp(a) on CVD, we have been unable to achieve either harmonization or standardization owing to its molecular heterogeneity8).
Lp(a) is a unique lipoprotein first discovered by Kare Berg in 1963 9). Lp(a) is a complex of LDL and apolipoprotein (a) [apo(a)], covalently linked by a disulfide bond. Apo(a) contains several tandem repeats of the plasminogen-like kringle IV (KIV) domain, followed by sequences with high homology to kringle V and protease domains of plasminogen. KIV is further classified into 10 types (designated KIV1 to KIV10). KIV2 (type 2 of KIV) exhibits an extremely wide diversity in molecular size because the number of KIV2 repeats ranges from 1 to >40, which is genetically determined by the LPA gene. Lp(a) concentrations differ across racial groups: the highest in Africans and the lowest in Asians10). The majority of the population expresses two apo(a) isoforms. Currently, the Lp(a) concentrations are measured by immunoassays using antibodies against apo(a)11). Some assay kits use polyclonal antibodies, while others use monoclonal antibodies. It is very likely that the differences in the Lp(a) epitopes recognized by these antibodies affected the measured results.
The International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) Working Group for Lp(a) Assay Standardization is establishing a reference measurement system for Lp(a) consisting of primary and secondary reference materials and a mass spectrometry (MS)-based reference measurement procedure (RMP). This RMP is endorsed by the IFCC (IFCC-MS-RMP) and expresses results in SI units (nmol/L)12). Until reference materials are established, the IFCC-MS-RMP is calibrated using the WHO/IFCC international reference reagent material SRM2B13). The commutability of the WHO/IFCC SRM2B has not yet been established11).
The purpose of the present study was to investigate whether we could harmonize Lp(a) measurements with different immunoassays using a serum panel value assigned with the IFCC-MS-RMP12).
Frozen serum samples collected from two institutions were used. Five hundred serum samples were randomly collected from the participants of the Bunkyo Healthy Study, a prospective cohort study conducted over 10 years14). This cohort consisted of approximately 1,000 elderly Japanese residents aged over 65 years in an urban area of Tokyo (Bunkyo-ku). Detailed background information has been described elsewhere14). The second set of serum samples was obtained from volunteers with diverse ethnic backgrounds by the Centers for Disease Control and Prevention (CDC), in Atlanta, GA, the United States of America (U.S.A.). Single donor samples were collected and processed according to CLSI C37-A protocol, as described by Danilenko et al.15) These unmodified serum samples had Lp(a) values assigned via the calibration laboratory in Leiden using the IFCC-MS RMP12).
Written informed consent was obtained from the blood donors at the CDC and from participants of the Bunkyo Health Study at Juntendo University. The serum was separated by centrifugation, aliquoted into multiple polypropylene screw-capped tubes, and frozen until use. At the time of enrollment in the Bunkyo Health Study, we did not include Lp(a) in the list of measurements, because the Lp(a) assays were neither harmonized nor standardized. In 2023, we began working with the IFCC and CDC on an Lp(a) harmonization/standardization project. Therefore, we decided to use stored serum for Lp(a) measurements using the opt-out method. The Institutional Review Board of Juntendo University (initial approval no. 2015078 and latest revised version no. M15-0057-M08), we posted the protocol and contact information of this study on the website of Juntendo University so that participants of the Bunkyo Health Study could request us not to use their serum samples for this study at any time. The study protocol was reviewed and approved by each institution and conducted in accordance with the tenets of the Declaration of Helsinki16).
3. 2. Lp(a) IFCC-MS RMP MethodTo guarantee that commercial Lp(a) tests produce accurate results, a more recently established “IFCC Working Group on Apolipoprotein Standardization by Mass Spectrometry” developed an MS-based RMP for apo(a)12). The RMP was operational in a network of three Calibration Laboratories. The sera used in the CDC panel were assigned by the Leiden Apolipoprotein Calibration Laboratory, Leiden, Netherlands.
3. 3. Lp(a) ImmunoassaysWe invited the manufacturers of Lp(a) assay kits to participate in this study through the Japan Association of Clinical Chemistry (JSCC) and the Japan Association of Clinical Reagents Industries (JACRI). Four manufacturers (Sekisui Medical, Denka, Shino-test, and Nittobo) accepted the invitation. Denka submitted two formulas: Denka-1 is the formula for the Japanese market, while Denka-2 is the formula for the foreign market. Although Denka-1 and Denka-2 use the same antibody for both kits, there are some differences in their formulas. The Shino-test submitted its prototype, which was not commercially available. We also invited Roche to participate in our study because the Lp(a) values determined using Roche’s reagent were shown to be traceable to those based on SRM2B. The Lp(a) values of the other reagents were determined using in-house standards. Of the five reagents, the Sekisui and Shino-test used monoclonal antibodies, while the others used polyclonal antibodies. The principle of the assay is turbidimetric immunoassay (TIA) for the Nittobo reagent and latex turbidimetric immunoassay (LTIA) for the others. The Roche reagent was run on Cobas pure, whereas the other reagents were run on LABOSPECT 008 (also called “Hitachi 917,” Hitachi High-Tech, Tokyo, Japan). Lp(a) results are reported in SI units (nmol/L) for the Roche reagent and in mass units (mg/dL) for the other reagents. In this study, “Lp(a)-X” was used to represent the Lp(a) value measured using Company X’s reagent. When the Lp(a) results exceeded the upper limit of the analytical range, the sample volume was automatically reduced or the samples were automatically diluted with saline. The principles and antibodies used for each kit are summarized in Supplemental Table 1.
Reagent (IVD company) |
A (Sekisui Medical) |
B (Denka-1) |
C (Denka-2) |
D (Shino-test) |
E (Nittobo Medical) |
F (Roche) |
---|---|---|---|---|---|---|
Brand Name |
Lp(a) Latex 「DAIICHI」 |
Lp(a)-LATEX “SEIKEN” | Lp(a) II-LATEX “SEIKEN” | Shino-test Lp(a) reagent (Prototype) | N-Assay TIA Lp(a) Nittobo | Tina-quant Lipoprotein (a) Gen.2 |
Principle | LTIA | LTIA | LTIA | LTIA | TIA | LTIA |
Type of antibody | Monoclonal | Polyclonal | Polyclonal | Monoclonal | Polyclonal | Polyclonal |
(Species) | (mouse) | (rabbit)#1 | (rabbit) #1 | (mouse) | (goat) | (rabbit) |
Calibrator | Lp(a)Latex Standard Serum L,M,H | Lp(a) Calibrators#2 (Made by Denka) | Lp(a) Calibrators#2 (Made by Denka) | Lp(a) Calibrators#3 (Made by Shino-test) | TIA Lp(a) Calibrator (Multi-point) | Preciset Lp(a) Gen.2 |
Traceability | In-house standard | In-house standard | In-house standard | In-house standard | In-house standard | SRM 2B |
(Certifciation organization) | (IFCC/WHO) | |||||
Quality control material | Lp(a) Control Serum “DAIICHI” LOW&HIGH | Lp(a) Control N Lp(a) Control AN | Lp(a) Control N Lp(a) Control AN | Lp(a) Control | Immunoquest Lp(a)-I Immunoquest Lp(a)-II | PreciControl Lp(a) Gen.2 |
Analyzer | LABOSPECT 008 | LABOSPECT 008 | LABOSPECT 008 | LABOSPECT 008 | LABOSPECT 008 | Cobas pure c303 |
(Ultrasonic mixing) | (Ultrasonic mixing) | (Ultrasonic mixing) | (Ultrasonic mixing) | (Ultrasonic mixing) | (Ultrasonic mixing) | |
Sample Vol. (mL) | 2.0 | 3.2 | 1.8 | 1.7 | 7.0 | 1.5 |
Reagent Vol. (mL) R1/R2 | 120/40 | 120/80 | 120/30 | 100/33 | 140/14 | 100/25 |
Abs WL | ||||||
2nd/primary (nm) | -/600 | -/700 | -/700 | -/600 | 700/340 | 800/660 |
Assay Mode | 2-point end assay | 2-point end assay | 2-point end assay | 2-point end assay | 2-point end assay | 2-point end assay |
Calibration | Spline | Spline | Spline | Spline | Line-Graph | Spline |
Analytical measurement range | 1-100 mg/dL | 0.5-80 mg/dL | 0.5-80 mg/dL | 1.3-80 mg/dL | 2.0-100 mg/dL | 7-240 nmol/L |
Automatic re-test | >100mg/dL | >80 mg/dL | >80 mg/dL | >80 mg/dL | >100 mg/dL | >240 mg/dL |
Sample reduction | -25% | N/A | N/A | -12% (80-100 mg/dL) | -50% | N/A |
Sample dilution | N/A | 1:3 (saline) | 1:3 (saline) | 1:3 (saline, >100 mg/dL) | N/A | 1:3 (saline) |
LABOSPECT 008 is the brand name in Japan and is the same autoanalyzer sold in Europe and the U.S.A. under the brand name Hitachi 917.
IVD, in-vitro diagnotics; Vol, volume; Abs WL, absorbance wavelength
Each reagent was run on corresponding analyzers.
LTIA, latex turbidimetric immunoassay; TIA, turbidimetric immunoassay; N/A, not applicable.
#1 Denka-1 (for the Japanese market) and Denka-2 (for the foreign market) use the same antibody, although there are some differences in the formulas.
#2 Denka-1 and Denka-2 use the same calibrator made by Denka.
#3 The Shino-test reagent use their own calibrator.
Because the Roche reagent was used as a reference in the Bunkyo Health Study samples, we evaluated the limit of blank (LoB), limit of detection (LoD), and limit of quantitation (LoQ) of the Roche reagent in the preliminary experiments. In the first step, the Lp(a) value of the calibrator blank, which did not contain Lp(a), was determined 60 times. LoB was determined as the mean+1.645× standard deviation (SD) using the non-parametric method (Supplemental Table 2). In the second step, we selected eight serum samples that were expected to have low Lp(a) near the expected LoD and measured the Lp(a) concentrations 10 times each. The composite SD was calculated as the square root of the sum of the SD2 of the eight samples. The LoD was determined as LoB+1.63×(composite SD). In the third step, we collected 13 samples whose Lp(a) concentrations were close to the expected LoQ and measured their Lp(a) concentrations in duplicate for five days. For each sample, we determined the mean, SD and coefficient of variance (CV). We plotted the CV values against Lp(a) concentration and determined 2.2 nmol/L as the LoQ at a CV value of 20%.
Step 1#1 | Step 2#2 | Step 3 | |
---|---|---|---|
Sample | Calibrator blank | ||
Total number of measurements | 60 | ||
Mean Lp(a) (nmol/L) | 0.1568 | ||
SD (nmol/L) | 0.2347 | ||
LoB | 0.5428 | ||
Sample (n) | Serum (n = 8) | ||
Total number of measurements | 80 (10 times each) | ||
Mean Lp(a) for all samples (nmol/L) | 2.646 | ||
Range of mean Lp(a) for each sample (nmol/L) | 1.670 − 4.235 | ||
Composite SD#1 (nmol/L) | 0.468 | ||
Range of SD for each sample (nmol/L) | 0.2723 − 1.012 | ||
LoD (nmol/L) | 1.3127 | ||
Sample (n) | Serum (n = 13) | ||
Total number of measurements | 130 (10 times each) | ||
Mean Lp(a) for all samples (nmol/L) | 6.660 | ||
Range of mean Lp(a) for each sample (nmol/L) | 0.976 – 23.18 | ||
Range of SD for each sample (nmol/L) | 0.272 – 1.012 | ||
Range of CV for each sample (%) | 1 .3 – 74.7 | ||
LoQ, according to CV | |||
CV 10% | 4.097 | ||
CV 20% | 2.220 | ||
CV 30% | 1.551 |
Lp(a) concentration was measured in calibrator blank containing no Lp(a) (Step 1), while it was measured in serum samples containing low levels of Lp(a) (Step 2 and Step 3).
#1 The composite SD was calculated as the square root of the sum of the SD2 of 8 samples. The SD value of each sample was calculated from Lp(a) values measured 10 times.
Lp(a) concentrations were measured in 40 serum panel samples collected at the CDC by participating immunoassays, including the Roche reagent. The Lp(a) concentrations in these samples were determined using the IFCC-MS-RMP with SI unit (nmol/L). Thus, the Lp(a) results are called the “Lp(a)-IFCC” and the detailed procedures of the IFCC-MS-RMP have already been published elsewhere12).
The Lp(a) results measured by the five tested reagents were plotted against Lp(a)-IFCC. Based on the relationship between Lp(a)-X and Lp(a)-IFCC, a linear regression equation was obtained for each reagent using the least-squares method. Using the inverse functions of the individual regression equations, Lp(a) results in the mass unit were converted to those corresponding to the SI unit. These converted Lp(a) results are designated as “Lp(a)IFCC-X.” The Lp(a) results obtained using the Roche reagent were converted in the same manner and designated as Lp(a)IFCC-Roche.
3. 5. Assessment of PrecisionThe precision of each assay was evaluated using both commercially available QC controls and homemade aliquoted pooled serum samples with low and high Lp(a) concentrations. Repeatability was determined by measuring Lp(a) concentrations in each sample 20 times within one day. CV was determined using the mean and SD. The intermediate precision was determined by measuring the Lp(a) concentration in each sample twice daily for 20 consecutive days in duplicate. The mean of the two measurements per day was used to calculate the CV. The total CV was calculated as the square root of (repeatability CV)2+(intermediate precision CV)2.
3. 6. Statistical AnalysisData are presented as the mean±SD. Because the Lp(a) concentrations were skewed to a lower value, the relationships between the different assay kits were analyzed using Spearman’s correlation coefficients. The means between the groups were compared using Student’s t-test. Statistical significance was set at p<0.05. These calculations were performed using the add-in software program Statcel (version 4.0) in Microsoft EXCEL.
All the tested reagents demonstrated satisfactory CV values for repeatability and intermediate precision. The total CVs was less than 3% for both the low and high Lp(a) QC controls (Table 1). In the pooled serum analysis, the total CVs ranged from 3 to 6% for the low Lp(a) serum, but were less than 3% for the high Lp(a) serum.
Reagent (Unit) | Sekisui (mg/dL) | Denka-1 (mg/dL) | Denka-2 (mg/dL) | Shino-test (mg/dL) | Nittobo (mg/dL) | Roche (nmol/L) | |
---|---|---|---|---|---|---|---|
QC material [low Lp(a)] | |||||||
Repeatability | |||||||
Mean* | 13.42 | 24.00 | 23.85 | 20.01 | 21.93 | 42.42 | |
SD | 0.14 | 0.30 | 0.31 | 0.19 | 0.29 | 0.49 | |
CV (%) | 1.06 | 1.26 | 1.28 | 0.93 | 1.32 | 1.17 | |
Intermediate precision | |||||||
Mean* | 12.66 | 24.36 | 25.51 | 19.34 | 21.47 | 41.57 | |
SD | 0.23 | 0.56 | 0.59 | 0.22 | 0.30 | 0.58 | |
CV (%) | 1.78 | 2.29 | 2.30 | 1.12 | 1.38 | 1.41 | |
Total CV (%) | 2.08 | 2.61 | 2.63 | 1.45 | 1.91 | 1.83 | |
QC material [high Lp(a)] | |||||||
Repeatability | |||||||
Mean* | 47.23 | 59.45 | 58.03 | 54.85 | 61.31 | 120.00 | |
SD | 0.41 | 0.30 | 0.37 | 0.53 | 0.38 | 0.65 | |
CV (%) | 0.86 | 0.50 | 0.64 | 0.96 | 0.62 | 0.54 | |
Intermediate precision | |||||||
Mean* | 46.13 | 57.41 | 57.80 | 52.22 | 60.54 | 116.80 | |
SD | 0.43 | 0.82 | 0.73 | 0.63 | 0.54 | 2.46 | |
CV (%) | 0.94 | 1.42 | 1.26 | 1.21 | 0.90 | 2.11 | |
Total CV (%) | 1.27 | 1.51 | 1.41 | 1.55 | 1.09 | 2.18 | |
Pooled serum 1 (low Lp(a)) | |||||||
Repeatability | |||||||
Mean* | 10.93 | 14.71 | 14.53 | 16.46 | 16.60 | 27.22 | |
SD | 0.15 | 0.22 | 0.25 | 0.17 | 0.27 | 0.45 | |
CV (%) | 1.40 | 1.52 | 1.70 | 1.01 | 1.62 | 1.65 | |
Intermediate precision | |||||||
Mean* | 0.96 | 13.96 | 14.50 | 14.81 | 14.83 | 25.59 | |
SD | 0.36 | 0.48 | 0.62 | 0.85 | 0.52 | 0.85 | |
CV (%) | 3.74 | 3.47 | 4.27 | 5.75 | 3.52 | 3.31 | |
Total CV (%) | 3.99 | 3.79 | 4.60 | 5.84 | 3.87 | 3.70 | |
Pooled serum 2 (high Lp(a)) | |||||||
Repeatability | |||||||
Mean* | 26.30 | 36.53 | 37.81 | 38.75 | 31.44 | 71.23 | |
SD | 0.19 | 0.34 | 0.30 | 0.31 | 0.24 | 0.45 | |
CV (%) | 0.73 | 0.93 | 0.80 | 0.80 | 0.77 | 0.63 | |
Intermediate precision | |||||||
Mean* | 23.87 | 35.49 | 38.04 | 36.30 | 31.77 | 69.43 | |
SD | 0.59 | 0.46 | 0.34 | 0.65 | 0.37 | 0.80 | |
CV (%) | 2.47 | 1.31 | 0.90 | 1.79 | 1.17 | 1.15 | |
Total CV (%) | 2.58 | 1.61 | 1.21 | 1.96 | 1.41 | 1.32 |
*The mean values of Lp(a) are presented in nmol/L for Roche reagent and in mg/dL for the other immunoassays.
The experimental protocol for the determination of the CV for repeatability and intermediate precision is described in detail in the Methods section. The total CV was calculated as the square root of (repeatability CV)2 + (intermediate precision CV)2.
Lp(a) concentrations measured by the tested immunoassays, including the Roche reagent, showed a good correlation with those measured by the IFCC-MS-RMP (rs>0.981) (Table 2). It should be noted that Lp(a)-IFCC and Lp(a)-Roche were traceable to WHO/IFCC SRM2B, whereas the Lp(a) values of the other reagents were calibrated to each company’s in-house working calibrators. The median Lp(a) values [25th - 75th percentile] measured by the IFCC-MS-RMP and Roche reagent were 56.6 [19.2 - 108.5] nmol/L and 50.7 [17.1 - 100.9] nmol/L, respectively. The correlations were better in the LTIA than in the TIA (Fig.1). Although all the regression lines ran approximately through the origin, their slopes varied widely. The slope for the Denka-1 reagent was almost twice that of the Sekisui reagent. The residual plots clearly show that Lp(a) values in the low and mid-range were more accurately predicted by these linear regression models than those in the high range (Fig.2). We obtained formulas that could convert the Lp(a) values in the X reagent to those in the IFCC-MS-RMP (nmol/L).
Reagent (Unit) |
Sekisui (mg/dL) |
Denka-1 (mg/dL) |
Denka-2 (mg/dL) |
Shino-test (mg/dL) |
Nittobo (mg/dL) |
Roche (nmol/L) |
---|---|---|---|---|---|---|
rs | 0.986 | 0.994 | 0.992 | 0.992 | 0.981 | 0.992 |
p-value | <0.000001 | <0.000001 | <0.000001 | <0.000001 | <0.000001 | <0.000001 |
Slope | 0.252 | 0.475 | 0.469 | 0.397 | 0.375 | 0.818 |
Constant | 1.64 | 2.47 | 2.19 | 2.53 | 6.72 | 4.87 |
Equation#1 to convert Lp(a)-X to Lp(a)IFCC-X | = 3.77*[C]−2.39 | = 2.04*[C]−2.77 | = 2.08*[C]−2.73 | = 2.48*[C]−5.01 | = 2.40*[C]−8.64 | = 1.16*[C]−1.94 |
Lp(a)-X, Lp(a) concentration measured with reagent X; Lp(a)-IFCC, Lp(a) concentration measured with IFCC-MS-RMP (IFCC-endorsed mass spectrometry-based reference measurement procedure) at Leiden University. The samples were collected from the CDC (n = 40).
Lp(a) concentrations were expressed in nmol/L (CDC and Roche) or in mg/dL (the other reagents).
The Lp(a)-IFCC concentration was measured using the IFCC-MS-RMP, while the Lp(a)-X concentration was measured using individual immunoassays. Note that the Lp(a)-IFCC concentration is shown in nmol/L on the X-axis. Lp(a)-X concentration is presented in mg/dL (panels A-E) and nmol/L (F) on the Y-axis.
The regression lines and their 95% confidence intervals are presented as solid and dotted black lines, respectively. The prediction intervals are represented by solid blue lines. The blue dotted line indicates the upper limit of Lp(a) for which we did not need to reduce the sample volume or dilute it with saline.
The Lp(a)-X concentration is shown in mg/dL (panels A-E) or nmol/L (panel F) on the X-axis. The residual was calculated by subtracting the predicted Lp(a) value from the actual Lp(a) value using a regression equation. The blue dotted line indicates the upper limit of Lp(a) for which we did not need to reduce the sample volume or dilute it with saline.
After converting the Lp(a)-X values to Lp(a)IFCC-X values in CDC samples, we listed the samples that had a difference of more than 20 nmol/L in Lp(a)IFCC-X among the four LTIAs. Five of the six samples had the highest Lp(a)IFCC-X values from the first to the fifth (Table 3). For the other samples, the average Lp(a)IFCC-X difference was 7.6±4.7 nmol/L.
Reagent (Unit) |
Lp(a)-IFCC#1 (nmol/L) |
Lp(a)IFCC-X (Converted value) | ||||
---|---|---|---|---|---|---|
Sekisui (nmol/L) |
Denka-1 (nmol/L) |
Shino-test (nmol/L) |
Roche (nmol/L) |
Max minus Min among 4 LTIAs |
||
1st highest#1 | 246.0 | 175.8 | 204.2 | 242.3 | 175.9 | 70.2 |
2nd | 234.0 | 227.1 | 233.5 | 212.6 | 207.3 | 26.7 |
3rd | 194.0 | 223.3 | 220.8 | 170.3 | 227.1 | 56.8 |
4th | 180.0 | 162.6 | 155.0 | 184.3 | 169.0 | 34.5 |
5th | 178.0 | 194.7 | 209.5 | 199.5 | 193.4 | 31.5 |
14th | 94.5 | 118.9 | 105.3 | 110.1 | 113.7 | 24.4 |
Upper Lp(a)IFCC-X for direct measurement | N/A | 374.6 | 160.4 | 163.7 | 231.4 | 114.1 |
Lp(a)IFCC-X, Lp(a) concentration measured with reagent X converted to the Lp(a)-IFCC equivalent value using the equations described in Table 1. Lp(a)IFCC-X is expressed in nmol/L in all reagents.
#1 Data are presented in the highest order of Lp(a)–IFCC values.
We measured Lp(a) concentrations in 500 healthy elderly subjects using five immunoassays and examined the correlations of those determined by the Roche reagent with those of the other reagents. Lp(a) concentrations were skewed to lower values (Supplemental Table 3). The median Lp(a) values [25th–75th percentile] measured by the Roche reagent were 18.3 [7.8-39.3] nmol/L, much less than those in the CDC samples. Lp(a) concentrations measured by LTIA correlated well with those measured by the Roche reagent, with rs ≥ 0.986 (Fig.3, Table 4, without conversion). The Lp(a) values with the Nittobo reagent correlated well with those with the Roche reagent in the middle to low concentration range but showed significant variation in the high concentration range (Fig.3, panel E). The residual plots showed that the Lp(a) values in the low and middle ranges were more accurately predicted by these linear regression models than those in the high range (Fig.4).
Class interval (nmol/L) | Class center (nmol/L) | Frequency | Cumulative frequency | Relative frequency |
Cumulative relative frequency |
---|---|---|---|---|---|
0~<20 | 10 | 259 | 259 | 0.518 | 0.518 |
~<40 | 30 | 104 | 363 | 0.208 | 0.726 |
~<60 | 50 | 54 | 417 | 0.108 | 0.834 |
~<80 | 70 | 29 | 446 | 0.058 | 0.892 |
~<100 | 90 | 15 | 461 | 0.030 | 0.922 |
~<120 | 110 | 8 | 469 | 0.016 | 0.938 |
~<140 | 130 | 9 | 478 | 0.018 | 0.956 |
~<160 | 150 | 5 | 483 | 0.010 | 0.966 |
~<180 | 170 | 4 | 487 | 0.008 | 0.974 |
~<200 | 190 | 3 | 490 | 0.006 | 0.980 |
~<220 | 210 | 3 | 493 | 0.006 | 0.986 |
~<240 | 230 | 0 | 493 | 0.000 | 0.986 |
~<260 | 250 | 3 | 496 | 0.006 | 0.992 |
~<280 | 270 | 3 | 499 | 0.006 | 0.998 |
~<300 | 290 | 1 | 500 | 0.002 | 1.000 |
Lp(a) concentration was measured using the Roche’s TIA kit on a LABOSPECT 008. We used the raw data without conversion to Lp(a)IFCC-Lp(a) concentration.
Lp(a) concentrations were measured using latex turbidimetric immunoassays, except for the Nittobo reagent (turbidimetric immunoassay; panel E). Note that the Lp(a)-Roche concentration is shown in nmol/L on the X-axis. The Lp(a)-X concentration is shown in mg/dL on the Y-axis (panels A-E). The black and blue solid and dotted lines are the same as those in Fig. 1.
Reagent | Sekisui | Denka-1 | Denka-2 | Shino-test | Nittobo |
---|---|---|---|---|---|
Lp(a)-X (mg/dL) (without conversion) | |||||
rs | 0.992 | 0.998 | 0.998 | 0.998 | 0.986 |
p-value | <0.000001 | <0.000001 | <0.000001 | <0.000001 | <0.000001 |
Slope | 0.292 | 0.579 | 0.574 | 0.443 | 0.522 |
Constant | 0.30 | 1.03 | 1.18 | 0.76 | 2.96 |
Lp(a)IFCC-X (nmol/L) (with conversion) | |||||
rs | 0.992 | 0.998 | - | 0.994 | 0.986 |
p-value | <0.000001 | <0.000001 | <0.000001 | <0.000001 | <0.000001 |
Slope | 0.945 | 1.018 | - | 0.943 | 1.078 |
Constant | 0.56 | 1.32 | - | 1.32 | 0.56 |
Lp(a) concentration measured with reagent X [Lp(a)-X] was converted to Lp(a)-IFCC equivalent value [Lp(a)IFCC-X] using the equations described in Table 1.
The Lp(a)-X concentration is expressed in mg/dL for all reagents except for Roche. Lp(a)-Roche and Lp(a)IFCC-X concentrations were expressed in nmol/L.
The Lp(a)-X concentration is shown in mg/dL on the X-axis. The residual was calculated by subtracting the predicted Lp(a) value from the actual Lp(a) value using a regression equation. The blue dotted line indicates the upper limit of Lp(a) for which we did not need to reduce the sample volume or dilute it with saline.
Using the formulas described in Table 2, we converted the Lp(a) assay results of the Bunkyo Health Study samples to values equivalent to the Lp(a) concentrations determined by the IFCC-MS-RMP. Because we did not have the data directly measured by the IFCC-MS-RMP, the Lp(a)IFCC-X values (converted results) of the tested reagents were plotted on the y-axis against the Lp(a)IFCC-Roche of each sample. Because the Lp(a) values were almost the same for the Denka-1 and Denka-2 reagents [Fig.1, panels B and C; Fig.3, panels B and C], we only converted the Lp(a) values of the Denka-1 reagent. After conversion, the slopes and constants of these correlation lines were almost 1.0, and zero for all reagents (Table 4).
In the Bunkyo Health Study, we had some samples with high Lp(a)IFCC-X values that were higher than the highest Lp(a) concentration in the CDC panel serum (246 nmol/L). As expected, the Lp(a)IFCC-X values were consistent among the tested reagents (Fig.5, panel A), especially among the LTIAs (Fig.5, panel B). The difference among LTIAs was acceptable, at least up to the 95 percentile Lp(a)IFCC-Roche (138 nmol/L). We then calculated the inter-method CV for each sample after excluding data where the original Lp(a) concentration was below the lower limit of the analytical range or where the converted Lp(a) values were negative for any reagent. They were significantly reduced after conversion, especially in the medium and high Lp(a) ranges (Fig.6). This effect was dominant in LTIA.
Lp(a) concentration was measured using five immunoassays. All Lp(a) results were converted to Lp(a)-IFCC equivalents [Lp(a)IFCC-X] in nmol/L, using the equations described in Table 2. Data from three latex turbidimetric immunoassays (LTIA) and one turbidimetric immunoassay (TIA) are shown in Panel A, and only those from three LTIA are shown in Panel B. The dotted lines indicate the upper limits for which we did not need to reduce the sample volume or dilute it with saline.
We used the raw Lp(a) results [Lp(a)-X] in mg/dL measured by each immunoassay (panel A) and the converted Lp(a) results [Lp(a)IFCC-X] in nmol/L (panel B). The intermethod CV value was calculated for each sample and plotted against the mean Lp(a) values. Samples were excluded from this analysis if the original Lp(a)-X values were lower than the lower limit of any reagent or if the converted Lp(a) IFCC-X values were negative.
We converted 30 mg/dL and 50 mg/dL in each reagent (except the Roche reagent) to Lp(a)IFCC-X values because these values are commonly used as cutoff values for diagnosing high Lp(a) concentrations in the national guidelines. Surprisingly, Lp(a) 30 mg/dL, determined by the Sekisui reagent, corresponded to approximately 50 mg/dL, as determined by other immunoassays (Table 5). To divide the subjects into low, intermediate, and high Lp(a) groups, we set tentative cutoff points at 50 and 100 nmol/L. The overall CV was calculated as the mean of the individual intermethod CVs of each group or all subjects. Before conversion, the overall CV values were similar among the three groups. After conversion, they decreased significantly in all Lp(a) groups. This was more pronounced in the medium- and high-Lp (a) groups than in the low-Lp (a) group (Table 6).
Cuoff value | Lp(a)IFCC-X values | ||||
---|---|---|---|---|---|
Sekisui | Denka-1 | Denka-2 | Shino-test | Nittobo | |
30 mg/dL | 112 | 58 | 59 | 69 | 62 |
50 mg/dL | 192 | 100 | 102 | 120 | 115 |
Values are expressed in nmol/L.
For each immunoassay, 30 mg/dL and 50 mg/dL were converted to Lp(a)IFCC-X values using the regression equation shown in Table 1.
Group Lp(a)IFCC-Roche (nmol/L) |
Low <50 (n = 324) |
Intermediate 50 – 100 (n = 57) |
High>100 (n = 32) |
All subjects (n = 413) |
---|---|---|---|---|
Overall CV (%) (LTIA + TIA) | ||||
Before conversion | 30.1 (6.9) | 27.5 (5.6) | 24.9 (4.2) | 29.2 (6.7) |
After conversion | 19.0 (15.3) | 12.2 (7.0) | 9.8 (5.3) | 17.1 (14.0) |
p-value | <0.000001 | <0.000001 | <0.000001 | <0.000001 |
Overall CV (%) (LTIA) | ||||
Before conversion | 29.1 (4.7) | 28.6 (4.2) | 26.6 (3.8) | 28.8 (4.6) |
After conversion | 15.2 (13.4) | 7.8 (3.8) | 6.2 (3.9) | 13.2 (12.3) |
p-value | <0.000001 | <0.000001 | <0.000001 | <0.000001 |
The subjects were divided into three groups based on Lp(a)IFCC-Roche 50 and 100 nmol/L.
The overall CV was defined as the average (SD) of the intermethod CVs of the individual samples.
LTIA, latex turbidimetric immunoassay (Sekisui, Denka-1, Shino-test); TIA, turbidimetric immunoassay (Nittobo).
The present study demonstrates that we can harmonize Lp(a) measurements with different immunoassays using a serum panel value assigned with the IFCC-MS-RMP. We found that the Lp(a)-X values in both CDC panel sera (Fig.1) and Bunkyo Health Study samples (Fig.3) showed a large variation among the immunoassays. After we converted Lp(a)-X concentrations to Lp(a)-IFCC equivalents [Lp(a)IFCC-X], the variation in Lp(a) values was significantly reduced, especially for the LTIA data (Fig.5 and 6).
Most international guidelines describe the importance of Lp(a) as a risk factor for CVD17-20), and some guidelines recommend that Lp(a) should be measured in all adults at least once in their lifetime for accurate risk assessment and CVD stratification19, 20). They also suggested a tentative clinical decision limit (CDL) for diagnosing a high Lp(a) concentration. The guidelines use 30 mg/dL or 50 mg/dL18-20). However, these CDLs have a potential risk of overestimation or underestimation owing to the lack of globally standardized methods for measuring Lp(a). In clinical practice, Lp(a) concentrations are measured by immunoassays using antibodies against apo(a). Some immunoassays use monoclonal antibodies, whereas others use polyclonal antibodies (Supplemental Table 1). Some experts are concerned about the accuracy of immunoassays that use polyclonal antibodies, because they may bind to KIV2, which has a large variation in the number of repeats8). If this is the case, it would be difficult to convert the Lp(a) concentrations determined by immunoassays to SI units. To overcome this situation, the IFCC established the first Lp(a) standardization working group in the 1990s, which developed a serum-based international reference material called IFCC SRM2B13). The Lp(a) concentration of SRM2B was assigned by an isoform-independent enzyme-linked immunoassay (ELISA) using two monoclonal antibodies developed by Marcovina et al.21). Although the excellent analytical performance of this ELISA has been confirmed using a large number of clinical samples22), global harmonization of Lp(a) has not yet been achieved11). According to the external quality assessment survey conducted in the Netherlands in 2018, the Lp(a) results still showed large interlaboratory variation. The highest Lp(a) values were almost twice as high as the lowest Lp(a) values11). Similar results were observed in the present study (Fig.1 and 3). Currently, SRM2B has been discontinued in Japan. The CDC announced that SRM2B was almost exhausted, and the ELISA-based KIV2-independent method was no longer available. Consequently a “new/2nd ” Working Group on Apolipoprotein Standardization was established in 2017 by IFCC, with the objective to meet the metrology gaps that had arisen (i.e., not only for Lp(a) but also for ApoA-I and ApoB standardization). The IFCC working group members decided to opt for SI-traceability of test results, that is, the highest order of calibration hierarchy, as explained in ISO 17511:2021, has to be developed. This was considered key for global Lp(a)/apo(a) standardization. Although other isoform-independent ELISAs using monoclonal antibodies23, 24) have been reported in the literature, their suitability as candidate RMP remains unknown.
Global standardization of serum lipids has been pursued by the CDC in collaboration with the Cholesterol Reference Method Laboratory Network (CRMLN). The CRMLN consists of highly specialized calibration laboratories worldwide and was established in 1988. The IFCC working group is closely working with the CDC to standardize Lp(a) in due time using the same approaches successfully applied in the CDC/CRMLN for traditional lipids. In this study, both the CDC CRMLN network coordinator and the IFCC APO MS WG chair facilitated this Lp(a) harmonization initiative by making available the 40 RMP value-assigned serum samples to manufacturers and laboratories. The present study demonstrates that panel sera with Lp(a) values assigned by IFCC-MS-RMP work very well to reduce the difference in reactivity of antibodies to Lp(a) between immunoassays (Fig.5 and 6, Table 6). Although some researchers are skeptical about achieving harmonization of Lp(a) immunoassays that use polyclonal antibodies, our conversion method works well even with such assays, especially around the Lp(a) cutoff values for the diagnosis or management of hyper-Lp(a)-emia. It is likely that manufacturers using polyclonal antibodies assigned Lp(a) values with different multiple calibrators of their own, which were not disclosed to the users.
Recent advances in Lp(a)-lowering therapy have increased the interest in the global standardization of Lp(a) measurements. Statins, the most effective oral LDL-C-lowering agents, increase rather than decrease Lp(a) concentrations25). Although other lipid-lowering agents such as niacin, microsomal triglyceride transfer protein (MTP) inhibitors, anti-PCSK9 antibodies, and cholesteryl ester transfer protein (CETP) inhibitors reduce Lp(a) concentrations, the maximum reduction rate is only approximately 40%26). In contrast, a daily dose of 600 mg SLN360, a short interfering RNA (siRNA) targeting hepatic apo(a) synthesis, reduced Lp(a) levels by more than 90% of basal levels in the Phase 1 APOLLO trial. The Lp(a) reduction was sustained for at least 150 days27). Another new class of drugs, pelacarcen, an antisense oligonucleotide (ASO), achieved more than 80% Lp(a) reduction at doses of 80 and 120 mg daily28). The German Apheresis Registry study reported the effects of lipoprotein apheresis on CVD events in 1,430 patients with established or progressive CVD. The median LDL-C and Lp(a) levels reduced to 67.5% and 71.1%, respectively. Both major coronary events (MACE) and major noncoronary events (MANCE) were reduced by more than 70% at one year29). Therefore, Lp(a) harmonization/standardization should be completed before these new therapies can be used in everyday clinical practice.
We must emphasize that we should advance the global Lp(a) standardization along with the transition from the mass unit (mg/dL) to the SI unit (nmol/L) for the following reasons. First, it is metrologically correct to express the Lp(a) concentration in the SI unit because of the heterogeneity in the molecular weight of Lp(a). The molecular weight of Lp(a) was 300–800 kDa, as determined by sodium dodecyl sulfate-polyacrylamide gel electrophoresis (SDS-PAGE). Similarly, the estimated molecular weight of Lp(a) from the sequence of apo(a) was 240–880 kDa30). Therefore, it is impossible to accurately express the Lp(a) particle concentration in mass units because of the different molecular weights of apo(a) and the variable lipid/protein composition of the particles. Second, we can implement Lp(a) values measured by different immunoassays at almost the same time during a certain pre-agreed transition period. The former success stories were the global IFCC HbA1c and IFCC enzyme standardization programs. In general, many clinicians are resistant to changes in laboratory test results, unless it is very well explained that mass results are scientifically flawed and not fit-for-clinical-purpose. Recently, the Japanese RMP for alkaline phosphatase (ALP) has changed from the JSCC method to the IFCC method within one year. The ALP (IFCC) values were almost one-third of the ALP (JSCC) values31), but there was little confusion throughout Japan. The case of Lp(a) is much simpler than that of ALP because Lp(a) is measured by far fewer doctors than ALP in Japan. In Western Europe, routine enzyme tests in medical laboratories generate test results that are fully traceable to IFCC Reference Measurement Systems within allowable measurement uncertainty. This is a matter of leadership and education of clinicians by laboratory professionals.
We must be aware that the present study is associated with some limitations. We evaluated only 5 Lp(a) immunoassays. One was a TIA kit using a polyclonal antibody, whereas the others were four LTIA kits using monoclonal (n=2) or polyclonal (n=2) antibodies. It appears that there is less variation in LTIA kits that use monoclonal antibodies, although this is not conclusive. To overcome this weakness, we are planning a future study to examine a larger number of Lp(a) reagents available not only in Japan, but also in the world market. With such data, we may be able to suggest preliminary CDLs for the diagnosis and management of elevated Lp(a) concentrations.
In conclusion, the present study showed that the harmonization of Lp(a) measurements among different immunoassays can be achieved using a serum panel value assigned with the IFCC-MS-RMP12). Second, laboratory professionals need to advocate proper implementation of the concept of metrological traceability of test results, especially for heterogeneous protein measurands, such as apo (a) in Lp(a). Setting a first step by harmonizing routine Lp(a) methods with the former WHO-IFCC reference material and expressing results in molar units would already harmonize test results among laboratories. Third, laboratory professionals should communicate about the relevance of unconfounded measurements and educate IVD-manufacturers, the pharmaceutical industry, and physicians on the clinical and metrology need for ultimate global standardization of Lp(a).
The findings and conclusions in this report are those of the authors (s) and do not necessarily represent the official position of the Centers for Disease Control and Prevention and the Agency for Toxic Substances and Disease Registry. The use of trade names is for identification only and does not imply endorsement by the Centers for Disease Control and Prevention, the Public Health Service, or the US Department of Health and Human Services.
This study was supported by the MEXT-Supported Program for the Strategic Research Foundation at Private Universities, 2014-2018 (S1411006) from the Ministry of Education, Culture, Sports, Science and Technology, and by the Japan Society for the Promotion of Science (JSPS) KAKENHI (Grant No. 15K08624 to T.M. and No. 18K07477 to S.H.) from the MEXT of Japan. We would like to thank all participants and staff of the Bunkyo Health Study.
The authors received Lp(a) immunoassay kits from Roche, Sekisui-Medical, Denka, Shino-test, and Nittobo. T.M. received honoraria for lectures from Sekisui-Medical, Denka, and Shino-tests.
T.M., S.H., Y.F., and A.H. designed the study, performed statistical analyses, and wrote the manuscript. S.I., M.H., and M.W. measured Lp(a) concentrations using an immunoassay. H.T., Y.T., H.W., and R.K. designed the Bunkyo Health Study, performed the blood sampling, and discussed the data. H.W.V. and C.M.C. provided advice on the study design and data analysis, and supervised the entire study. All the authors approved the final version of the manuscript.