1984 Volume 35 Issue 5 Pages 361-365
Ignition or combustion of endotracheal tube during CO2 laser laryngomicrosurgery may cause fatal lung disorders. We had one experiment on ignition and combustion of endotracheal tubes by CO2 laser beam. And We obtained several conclusions as follows.
(a) Silicon endotracheal tube was non-flammable.
(b) Spiral endo-tracheal tube was also non-flammable.
(c) Pulse wave was recommended and three second interval was necessary.
(d) Four sheets of moist gauze were necessary to prevent cuff-trouble.
(e) It was desired to set O2 concentration below 25% in the use of silicon endotracheal tube.