The Journal of Japanese College of Angiology
Online ISSN : 1880-8840
Print ISSN : 0387-1126
ISSN-L : 0387-1126
Original Article
Mechanical Evaluation of Stenting for Saccular Abdominal Aortic Aneurysm Using Fluid Structure Interaction Analysis
Masanori MurakamiFei JiangShahreen Bin Saiful YewXian Chen
Author information
JOURNAL OPEN ACCESS

2022 Volume 62 Issue 11 Pages 111-119

Details
Abstract

One of the factors that contribute to the development of aortic aneurysms is hemodynamic factors. In this study, we analyzed the hemodynamic changes in a stented model of a saccular abdominal aortic aneurysm to improve hemodynamic factors. We discovered that stenting reduced wall shear stress, mean flow velocity in the aneurysm, energy loss, and pressure loss coefficient, and improved hemodynamics around the saccular abdominal aortic aneurysm. Stenting is expected to become a new treatment modality for abdominal aortic aneurysms.

Content from these authors

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。
https://creativecommons.org/licenses/by-nc-nd/4.0/deed.ja
Next article
feedback
Top