Environmental Monitoring Method of Di-, Tri-, Tetra-, Penta-, and Hexa-Chlorinated Biphenyls in the Gas Phase by Use of the Picosecond Laser Ionization
Shinsaku DobashiYoshiki YamaguchiYoshinori IzawaYoshihiro DeguchiAkihide WadaMichikazu Hara
Author information
JOURNAL RESTRICTED ACCESS

2007 Volume 40 Issue 2 Pages 191-197

Details
Abstract

Utilization of the picosecond (100 ps) laser ionization was successful in measuring the mass spectra attributed to di-, tri-, tetra-, penta-, and hexa-chlorinated biphenyls. It was confirmed that the quantitative limit of this method in real-time (1 min) was 10 μg/m3N (equivalent to 1/10 of the value established as the statutory limit for PCBs in the ambient air of the workplace environment). Further evaluations of measurement accuracy based on comparisons with the results of the conventional method (gas sampling and GC-MS) demonstrated a coefficient of variation within 30%. We undertook the measurement of the exhaust gas and the work environment atmosphere in PCBs treatment plants using the LI-IT-TOFMS method, confirming that the monitoring system could be operated appropriately by the introduction of an internal standard gas (2,4-dichlorinated toluene), and that it was possible to make continuous long-term measurements for over 500 h (total measurement time, >6000 h).