Fatigue Testing System and Result of Long-term Operation
Tetsumi YURIKotobu NAGAIToshio OGATAOsamu UMEZAWAKeisuke ISHIKAWA
Author information
JOURNAL FREE ACCESS

1991 Volume 26 Issue 3 Pages 184-189

Details
Abstract
A new type of cryogenic temperature fatigue testing machine was installed in 1983 at the National Research Institute for Metals, Tukuba Laboratories. This machine is featured by the continuous operation at 4K without replenishing liquid helium. A closed-loop helium recondensation system enables a long-term fatigue test at 4K longer than 1, 000h. The total operation time reached 5, 500h in these seven years.