The Journal of Japan Endodontic Association
Online ISSN : 2432-4493
Print ISSN : 1340-6248
Original Article
A Study on the Handpiece with Electronic Canal Length Measurement part 2. Load Applied to Files
Chihiro KOBAYASHITakatomo YOSHIOKAHideaki SUDAKazunari MATOBA
Author information
JOURNAL FREE ACCESS

1996 Volume 17 Issue 1 Pages 41-48

Details
Abstract

Abstract : We have developed and already reported on a new cordless engine driven root canal preparation system which electronically monitors the location of the file tip during instrumentation procedures.

 The load, torque, rotation speed, and current flowing through the motor were measured when this system was used with nickel-titanium files for the preparation of human extracted tooth canals.

 The maximum load applied to the tooth was approximately 800 gram. The maximum torque observed in this system was approximately 450 g-cm. There was a linear relation between the magnitude of the torque and the current flowing through the motor. It was assumed that the Auto-torque-reverse mechanism should start at 40-80 g-cm when the ProFile or McXIM file was used. This torque range corresponded to the motor current of 100-200 mA. The rotation speed of the motor decreased as the load or torque increased, although this change was in a small range (240-280 rpm).

 This Auto-torque-reverse mechanism promises to prevent file breakage.

Content from these authors
© 1996 Japan Endodontic Association
Previous article Next article
feedback
Top