Calibration of the Balloon-Borne NO Instrument
Yutaka KONDONoriji TORIYAMAW. Andrew MATTHEWSPatrick AIMEDIEU
Author information
JOURNAL FREE ACCESS

1989 Volume 41 Issue 6 Pages 507-523

Details
Abstract
The precision of balloon-borne NO measurements using chemiluminescent instruments critically depends on the precision in estimating the flow rates of sampled air and calibration gas, as well as the concentration of NO in the calibration gas. These parameters have now been measured in the laboratory with more precision than was done for our previous work. The NO data we obtained from balloon observations in the past has been revised in light of these new laboratory calibration data.