Japanese Journal of Health Physics
Online ISSN : 1884-7560
Print ISSN : 0367-6110
ISSN-L : 0367-6110
Calibration of Ionization Chamber and GM Counter Survey Meters (I)
Tests for Measuring the β-ray Absorbed Dose Rate in a Working Area
Kazuyoshi BINGOYoichi KAJIMOTOShin-ichi SUGA
Author information
JOURNAL FREE ACCESS

1978 Volume 13 Issue 4 Pages 279-285

Details
Abstract

Three types of ionization chamber survey meters and a type of GM counter survey meter were calibrated for measuring the β-ray absorbed dose rate in a working area. To estimate the β-ray absorbed dose rate, a survey meter was used without and with a filter. A reading of survey meter's indicator measured with the filter was subtracted from a reading measured without the filter, and then the absorbed dose rate was obtained by multiplying this remainder by a conversion coefficient.
The conversion coefficients were roughly constant with distance more than 8cm (ionization chamber survey meters) and with distance more than 5cm (GM counter survey meter). The conversion coefficient was dependent on β-ray energies. In order to measure the absorbed dose rate of tissue whose epidermal thickness is 40mg/cm2, the constant value, 4(mrad/h)/(mR/h), was chosen independently of β-ray energies as the conversion coefficient of three types of ionization chamber survey meters. The conversion coefficient of the GM counter survey meter was more energy dependent than that of every type of ionization chamber survey meter.

Content from these authors
© Japan Health Physics Society
Previous article Next article
feedback
Top