Japanese Journal of Health Physics
Online ISSN : 1884-7560
Print ISSN : 0367-6110
ISSN-L : 0367-6110
A New Method for Evaluating Weak Leakage Gamma-Ray Dose Using a 3″φ×3″ NaI(Tl) Scintillation Spectrometer (III)
Expanding of Applicable Energy Range
Hideharu MATSUDAShigeko FURUKAWATokishi KAMINISHISusumu MINATO
Author information
JOURNAL FREE ACCESS

1987 Volume 22 Issue 3 Pages 283-286

Details
Abstract

Applying the leakage gamma-ray which energy exceeding primary gamma-ray of K-40, namely 1.46MeV, a method previously reported, this is, three component method is still not satisfactory. Therefore, a better approach for expanding applicable energy range was examined. The result of measurements showed that there was sufficiently a correlation between background exposure rate and concentration of Th. The present method by background exposure rate was estimated from only concentration of Th contained in environment was practically usable technique. Though the measured error will increase, applicable energy range of three component method could expanded to about 2.5MeV by this technique.

Content from these authors
© Japan Health Physics Society
Previous article Next article
feedback