Present Status and Problem of the Personnel Neutron Monitoring and Dosimeter Calibration
Takashi NAKAMURA
Author information
JOURNAL FREE ACCESS

1991 Volume 26 Issue 4 Pages 367-382

Details
Abstract
The present situation of personnel neutron monitoring is summarized in the facilities where the neutron exposure really happens, such as reactor, accelerator and nuclear fuel handling facilities, and the actual calibration method using a phantom is described for personnel neutron dosimeters (film badge, albedo TLD, solid state track detector) of practical use in these facilities.
The calibration method of neutron dosimeters is not standardized at present, and it is usually carried out using 252Cf and 241Am-Be neutron sources. Neither the energy spectra of these neutron sources are similar to those of actual neutron fields in the facilities nor are the energy responses of these neutron dosimeters coincident with the fluence-to-dose equivalent conversion factor defined by ICRP; it is thus difficult to evaluate the neutron dose equivalent from the measured data. Here, we refer to some typical neutron spectral fields and propose to use the moderating radioactive neutron source field, which has a similar spectrum to that of the actual field, for calibration of personnel neutron dosimeter.