Japanese Journal of Health Physics
Online ISSN : 1884-7560
Print ISSN : 0367-6110
ISSN-L : 0367-6110
Measurement of 131I in the Human Thyroid Gland with a Collimated Spectrosurvey Meter for the Case of Emergency Exposure
Mitsuru NENOIKiyoko IMAIHisamasa JOSHIMA
Author information
JOURNAL FREE ACCESS

1996 Volume 31 Issue 4 Pages 443-445

Details
Abstract
Feasibility for the use of the collimated NaI (Tl) scintillation spectrosurvey meter SS-γ C3475 (Hamamatsu Photonics) in the estimation of thyroidal 131I burden in patients that received emergency exposure to radioiodine (131I) is demonstrated. When the NaI (Tl) crystal probe collimated well to a neck phantom simulating the human thyroid gland, the detection limit of 131I deposited in that organ was estimated at 0.58kBq, which was well below the generally adopted screening level of 30kBq. The detection system for thyroidal 131I presented in this report is remarkably inexpensive compared to the conventional whole-body counters, and is considered to provide a satisfactorily accurate evaluation of thyroidal 131I burden.
Content from these authors
© Japan Health Physics Society
Previous article Next article
feedback
Top