Kenichi IijimaMasaomi ChinushiAkinori SatoYukio HosakaDaisuke IzumiHiroshi FurushimaHiroshi WatanabeMasahito SatoKatsuya EbeHiroshi ShimizuKazuyoshi TakahashiYoshifusa Aizawa
Author information
Keywords: lead trouble, ICD
JOURNAL FREE ACCESS

2011 Volume 27 Issue Supplement Pages OP53_2

Details
Abstract
Method: We investigated the outcomes of 136 patients with Sprint-Fidelis in 5 centers in Niigata Prefecture. Results Incidence: Lead troubles occurred in 9 patients (6.6%). The mean period before the trouble was 31 months (16–57 months). Implantation: Among these 9 patients, the lead was implanted using a cut-down technique in 4, an extra-thoracic approach in 4 and a subclavian approach in 1. Discovery: The lead troubles in 5 were discovered at regular clinics. The other 4 patients emergently visited because of the lead integrity alert (LIA) or inappropriate ICD discharges. In the 9 patients, the average sensing integrity counter was 338/day and mean F-F interval, of which was recognized as NST, was less than 200 ms. The LIA: The LIA had been programmed in 7, and 5 of the patients visited before inappropriate ICD discharges. However, the other 2 elderly patients couldn’t notice the alert sound. Extraction: Lead extraction was attempted in 6 and 5 leads were removable without any complications. Conclusions: The incidence of troubles with Sprint-Fidelis is high, and it occurs in the relatively late phase. The LIA is useful to detect the lead problems but its efficacy may be limited in elderly patients.
Content from these authors

This article cannot obtain the latest cited-by information.

© 2011 Japanese Heart Rhythm Society
Previous article Next article
feedback
Top