Journal of Arrhythmia
Online ISSN : 1883-2148
Print ISSN : 1880-4276
Reviews
Detection of atrial tachyarrhythmias in an implantable monitoring device
Haruhiko Abe
Author information
JOURNAL FREE ACCESS

2012 Volume 28 Issue 1 Pages 19-25

Details
Abstract
State-of-the-art, implantable, dual-chamber cardiac devices provide useful diagnostic information, including the number and duration of automatic-mode switch episodes in cases of atrial tachycardia and atrial fibrillation encountered in clinical practice. However, to acquire accurate diagnostic information, special attention must be paid to the device settings; to the presence or absence of ventriculoatrial conduction, which, when present, often represents repetitive non-reentrant synchrony (RNRVAS) or pacemaker-mediated tachycardia; to the post-ventricular atrial-blanking period and atrial sensitivity; and to the sensing of far-field R waves (FFRW) in the atrial channel. Physicians should be careful about the information gathered during the monitoring of patients with implantable devices.
Content from these authors

This article cannot obtain the latest cited-by information.

© 2012 Japanese Heart Rhythm Society