Accuracy evaluation of a non-invasive measuring instrument for diagnostic X-ray equipment
Hiroki SaitohShinji AbeIzumi OguraYoh KatohToru Negishi
Author information
JOURNAL FREE ACCESS

2014 Volume 17 Issue 1 Pages 43-50

Details
Abstract

The aim of this research is to assess the accuracy of the non-invasive measuring instrument, which is used for routine quality control of diagnostic X-ray equipment. We chose to measure the following parameters : X-ray tube voltage, half value layer (HVL), dose and irradiation time. Along with the precision of the mentioned instrument, we assessed X-ray tube voltage dependency based on various factors (X-ray tube current, irradiation time, SDD and added filtration). The instrument used to measure the X-ray tube voltage was the invasive X-ray tube voltage and current meter, and to measure the dose with ionization Chamber was used. The X-ray tube voltage error and the error in its dependence were within ±3%, this indicated high accuracy. The coefficient of determination R^2 was 0.99, indicating a good correlation. Based on the results, This non-invasive measuring instrument demonstrated high measurement accuracy, and therefore it is a reliable meter for routine quality control.