1993 Volume 86 Issue 6 Pages 875-879
We are using a new three-dimensional (3-D) surface reconstruction system to measure the temporal bones. This system offers the advantage of observation of the external aperture of the vestibular aqueduct and the porus acusticus internus in living subjects. However, its accuracy has not been confirmed. To investigate the accuracy of this new system, we measured the length of an in situ ceramic ossicular replacement prosthesis (CORP) of known length of 6.0mm using 3-D surface reconstruction, conventional plain X-ray and polytomography. The CORP was scanned in the axial, sagittal and oblique directions. The mean measured length obtained with the 3-D surface reconstruction images was 5.94±0.21 on vertical scans, 5.91±0.27 on horizontal scans, and 6.01±0.25 on oblique scans. There were no significant differences among the measured lengths obtained in the three directions. Therefore, this 3-D surface reconstruction measurement system is considered to be reliable.
Conversely, the mean measured length obtained by plain X-ray was 7.98±0.20, and by polytomography it was 7.94±0.23. These conventional methods have the inherent disadvantage of magnification of size which consequently requires correction.