Japanese Journal of Cardiovascular Surgery
Online ISSN : 1883-4108
Print ISSN : 0285-1474
ISSN-L : 0285-1474
A Rare Case of Primary Wireform Fracture Implanted in the Mitral Position of Carpentier-Edwards Pericardial Xenograft
Hideki YaoTakashi MiyamotoKatsuhiko YamashitaKazushige InoueHirokazu MinamimuraTorazou WadaHiroe TanakaMasaaki RyomotoYasuzumi HiraiTomohiko Sugimoto
Author information
JOURNAL FREE ACCESS

1998 Volume 27 Issue 2 Pages 125-128

Details
Abstract
A 71-year-old woman underwent mitral valve replacement with Carpentier-Edwards Pericardial Xenograft for mitral regurgitation on January 8, 1987. She had nocturnal hematuria and paroxysmal nocturnal hemoglobinuria was diagnosed in November, 1992. She had congestive heart failure in February, 1993. Cineradiographical analysis revealed a fracture of the wireform in three locations of the Xenograft and the stent was distorted inward. A second mitral valve replacement was successfully performed on March 16, 1993. She was discharged 45 days after operation after an uneventful course.
Content from these authors
© The Japanese Society for Cardiovascular Surgery
Previous article Next article
feedback