Japanese Journal of Drug Informatics
Online ISSN : 1883-423X
Print ISSN : 1345-1464
ISSN-L : 1345-1464
Short communication
Effectiveness of Countermeasures against Technical Alarms in the Dispensing Inspection Support System
Tomohiro UedaRika FukuiYuki KunitsuTakaki KamiyaDaiki HiraTomohiro TeradaShin-ya Morita
Author information
JOURNAL FREE ACCESS

2022 Volume 24 Issue 2 Pages 105-110

Details
Abstract

Objective: Our previous study reported many technical alarms in the dispensing inspection support system due to setting or operation errors, but not due to dispensing errors, and that countermeasures reduce technical alarms. In this study, we investigated the long-term effects of the countermeasures.

Methods: The log data of the inspection results were extracted from the dispensing inspection support system and divided into image inspection and weight inspection data. The details and frequency of technical alarms before (April-June 2019), after (April-June 2020), and one year after (April-June 2021) the countermeasures were compared by chi-square test.

Results: The frequency of technical alarms in image inspection was 13.7, 11.3, and 9.1% in 2019, 2020, and 2021, respectively (p< 0.01). However, the frequency of technical alarms in weight inspection was 3.0, 3.0, and 1.6% in 2019, 2020, and 2021,respectively (p< 0.01). The ratio of technical alarms to all alarms was 98.2% for image inspections and 91.2% for weight inspections in 2021.

Conclusion: The frequency of technical alarms was reduced by the countermeasures. However, the ratio of technical alarms to all alarms is still high, and further improvements are needed to prevent the loss of reliability due to technical alarms.

Content from these authors
© 2022 Japanese Society of Drug Informatics
Previous article Next article
feedback
Top