2025 Volume 61 Issue Supplement Pages 2F07-01
To improve movement accuracy, exoskeleton control must consider natural human motion. This study investigates how stiffness (i.e., the persistence of exoskeleton in following the reference trajectory) and pre-emptive offset (i.e., how far ahead in time) affect movement accuracy in an impedance control the exoskeleton that provides assistance preemptively based on a predicted trajectory. Thirteen participants performed an elbow flexion from 40° to 135° within 1.12 seconds while holding a load equivalent to 15% of their maximum voluntary contraction under exoskeleton assistance. The experiment tested three stiffness levels (0.1, 0.2, and 0.31 N m deg-1) and two pre-emptive offset levels (0.1 and 0.4 s). The results indicate that lower stiffness combined with a less aggressive pre-emptive move causes positional undershoot, whereas increasing both parameters causes overshoot, which intensifies with higher stiffness and a more aggressive pre-emptive move. To enhance movement accuracy, we suggest avoiding simultaneous increases or decreases in both parameters.