Japanese Journal of Extra-Corporeal Technology
Online ISSN : 1884-5452
Print ISSN : 0912-2664
ISSN-L : 0912-2664
Original Articles
Evaluation of spoilage of circulating water in a heater-cooler unit tank using adenosine triphosphate measurement: Practical water management method of a heater-cooler unit tank
Hitomi OkadaNaoki MomoseRie KusauraRyo KokuboChinori UmedaToru YasudaNorio IwamotoAtsushi Yamaguchi
Author information
JOURNAL FREE ACCESS

2017 Volume 44 Issue 2 Pages 88-93

Details
Abstract

A heater-cooler unit tank is indispensable for the heart-lung machine, however, spoilage of circulating water used there will threaten the patients’ life. Tap water collected from rivers and subjected to chlorine disinfection, but contains a trace amount of organic matter from soil and wastewater. This organic substance promotes the growth of bacteria and causes water spoilage. Reverse osmosis (RO) water is produced in large amount in the dialysis chamber. Not only organic matter but also electrolytes are removed by the filtration membrane during the production process of RO water, which suggests that it is unsuitable for bacterial growth.

In this study, we made a mock heater-cooler unit tank, evaluated water spoilage by adenosine triphosphate (ATP) determination in circulating water, and compared and examined depending on the type of circulating water used. Tap water±disinfectant and RO water±disinfectant were used for circulating water to be tested.

Samples using RO water showed no increase in ATP even after 28 days and were below 10 relative light unit(RLU). On the other hand, in samples using tap water, ATP after 28 days was significantly increased to a maximum of 1899 RLU. Even in visual observation, sediments were recognized regardless of whether or not disinfectants were used in the samples using tap water, which was cloudy. However, there were no sediments in the samples of RO water.

It is recommended to use RO water for circulating water in a heater-cooler unit tank in an operating room where cleanliness is required.

Content from these authors
© 2017 Japanese Journal of Extra-Corporeal Technology
Previous article
feedback