Japanese Journal of Medical Physics (Igakubutsuri)
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
ORIGINAL CONTRIBUTIONS
Development of the Image Registration Program for Portal and DRR Images in Radiation Therapy
Hiroyuki WatanabeTakeshi ItoKazuma NakazekoAtsuhi TachibanaTakeyuki HashimotoHiroyuki Shinohara
Author information
JOURNAL FREE ACCESS

2011 Volume 31 Issue 3 Pages 65-74

Details
Abstract
In this article, the authors propose an image registration program of portal images and digitally reconstructed radiography (DRR) images used as simulation images for external beam radiation therapy planning. First, the center of the radiation field in a portal image taken using a computed radiograhy cassette is matched to the center of the portal image. Then scale points projected on a DRR image and the portal image are deleted, and the portal image with the radiation field is extracted. Registration of the DRR and portal images is performed using mutual information as the registration criterion. It was found that the absolute displacement misregistrations in two directions (x, y) were 1.2±0.7 mm and 0.5±0.3 mm, respectively, and rotation disagreement about the z axis 0.3±0.3°. It was concluded the proposed method was applicable to image registration of portal and DRR images in radiation therapy.
Content from these authors
© 2011 Japan Society of Medical Physics
Previous article Next article
feedback