Japanese Journal of Medical Physics (Igakubutsuri)
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
REVIEWS
Beam Quality Conversion Factor
Toru KawachiTetsurou Katayose
Author information
JOURNAL FREE ACCESS

2013 Volume 33 Issue 4 Pages 186-191

Details
Abstract
This report describes the update of the beam quality conversion factor KQ,Q0 of the standard dosimetry protocol in Japan. The KQ,Q0 corrects for the difference between the response of an ionization chamber in the reference beam quality Q0 used for calibrating the chamber and in the actual user beam quality Q. All changes of KQ,Q0 were caused by the perturbation correction factors which were recalculated by Monte Carlo simulation. With a calculation process, unsolved problems in this update are also discussed here.
Content from these authors
© 2013 Japan Society of Medical Physics
Previous article Next article
feedback