Japanese Journal of Medical Physics
Online ISSN : 2186-9855
Print ISSN : 0918-8010
ISSN-L : 0918-8010
Estimated equation of Patient dose in Diagnostic radiology
Hiroki ShimazakiShinji InagakiHirotoshi KabeyaShoichi SuzukiShigehisa FujiiYasuki AsadaTakeo OritoSukehiko Koga
Author information
JOURNAL FREE ACCESS

1999 Volume 19 Issue 4 Pages 209-217

Details
Abstract

IAEA in Basic Safety Standards (1994) presented Guidance levels of dose for typical Diagnostic radiology. Then, IAEA advises us against being over this level. Again, ICRP in ICRP Publication 60published setting “dose constraint” as upper bounds of individual exposure.
However, many medical treatment facilities don't have a dosimet er in order to measure patient dose.
Then, we measured X-ray output of many X-ray equipments. We statistically analyzed many data measured and could make an approximate equation in order to estimate X-ray output (air ab sorbed dose) of X-ray equipments.
The X-ray output is directly proportional to V2.232, where V is the tube voltage. The coefficient of determination showed a high value.
By revising this approximate eq uation with various revision coefficients, we made an estimated equation could estimate patient dose in Diagnostic radiology. We think that this estimated equation of patient dose can easily employ any medical treatment facilities.

Content from these authors
© The Japan Society of Medical Physics
Next article
feedback