Japanese Journal of Medical Physics (Igakubutsuri)
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
Determination of overall perturbation factors forplane-parallel ionization chambers in electron beams
Fujio ArakiYuichi ShirakawaRyuji IkedaToshiaki ShimonobouNobuyuki MoribeTakao TakadaMutsumasa TakahashiHiroki OuraMasaru Matoba
Author information
JOURNAL FREE ACCESS

2000 Volume 20 Issue 2 Pages 95-103

Details
Abstract
Most dosimetry protocols recommend the use of plane-parallel chambers for dose determination in electron beams with energies below 10 MeV. The new IAEA TRS 381 (1997) protocol includes the overall perturbation factor PQ that consists of the inscattering correction factor Pcav(ar Prepi) and the wall correction factor Pwall(or Pwall)I. n this work, PQc for the commonly applied NACP, PTW/Roos and PTW/Markus planeparallel chambers was determined experimentally. For the NACP plane-parallel chamber, PQ was obtained by comparison with a cylindrical Farmer chamber, while for the PTW/Roos and PTW/Markus chambers it was obtained by comparison with the NACP chamber. The values of PQ for these plane-parallel chambers were measured as a function of mean electron energies ,Ez from 1.7 MeV to 11.5 MeV. It was found that for the NACP and PTW/Roos chambers, PQ is independent of energy down to Ez= 1.7MeV, while for the PTW/Markus chamber it shows a systematic and exponential drop of about 2% with decreasing energy down to Ez=2.7 MeV. However, the decrease of PQ for Ez=1.7 MeV was not exponential.
Content from these authors
© The Japan Society of Medical Physics
Previous article
feedback
Top