Japanese Journal of Medical Physics (Igakubutsuri)
Online ISSN : 2186-9634
Print ISSN : 1345-5354
ISSN-L : 1345-5354
Three-dimensional conformal radiotherapy for extracranial tumors using a stereotactic body frame.
Yasushi Nagata, Yoshiharu Negoro, Tetsuya Aoki, Takashi Mizowaki, Kenji Takayama, Masaki Kokubo, Norio Araki, Michihide Mitsumori, Sinsuke Yano, Yuko Koga, Masahiro Hiraoka
Author information
JOURNAL FREE ACCESS

2001 Volume 21 Issue 1 Pages 28-34

Details
Abstract
This paper was reviewed to evaluate the feasibility of three-dimensional (3-D) conformal radiotherapy for extracranial tumors, especially for solitary lung tumors using a stereotactic body frame.
To extend the te chnique of stereotactic irradiation for intracranial tumors, accurate body fixation and regulation of internal target motion are essential. In our study, a stereotactic body frame was used, and daily setup accuracy was verified. As a result, its setup accuracy was maintained within 0-8.5 mm (Ave = 2.5mm). In our initial clinical experiences for thirty-two patients with 6-10 non-coplanar static beams, forty or 48 Gy was irradiated. During the follow-up of 4-27 (Average =11) months, twentynine (94%) tumors were locally controlled without any symptomatic complications. Recently, respiratory-gated irradiation systems, CT-linac systems, a real-time tumor tracking system, Cyber-knife, and C-arm linac were developed. With all these techniques, stereotactic irradiation for extracranial tumors are future direction of threedimensional conformal radiotherapy.
Content from these authors
© The Japan Society of Medical Physics
Previous article Next article
feedback
Top