Dose reduction by several measures in X-ray angiographic system
Takaji Saito
Author information
JOURNAL FREE ACCESS

2002 Volume 22 Issue 2 Pages 105-109

Details
Abstract

“The radiation protection measures from equipment side, from a maker's point of view.”TAKAJI SAITO Medical Solutions Marketing Division, Siemens Asahi Medical Technologies Ltd.
Takin g care of limited fluoro and acquisition time during every case is an important prerequisite for dose reduction. Latest technology provides us with a number of means for dose savings regarding patients and staff.
Here, the practical techniques of the radiation control function in angiography system are addressed.
•Last Image Hold (LIH)
•Collimation without radiation on LIH
•Pulsed fluoroscopy at high and low rates
•Additional spectral filters in the X-ray beam
•Display of applied dose during the procedure
These technological measures will d efinitely contribute for dose reduction for interventional procedures.